HIGH PRESSURE ROTATOR WITH MALE LUER LOCK, List Number MX496HP
Published January 24, 2024 · Recall initiated: November 10, 2023
Summary
FDA records medical device Class II recall Z-0743-2024, published January 24, 2024 and initiated November 10, 2023. Recalling firm: Smiths Medical Asd Inc; Status: Ongoing; Reported quantity: 31,685 total; Lot: Lot Number: 4332915 4408772. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Smiths Medical Asd Inc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, UAE.
- Lot
- Lot Number: 4332915 4408772
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694RJanuary 24, 2024
- FDAHIGH PRESSURE ROTATOR WITH FEMALE LUER LOCK, List Number MX497HPJanuary 24, 2024
- FDA6IN PRESSURE TUBING, List Number MX20617January 24, 2024
- FDAMedex LOGICAL CATH LAB KIT, List Number M20754January 24, 2024
- FDAMedex 700PSI STOPCOCK W/ROTATOR, List Number MX4331RJanuary 24, 2024
- FDA20IN (50.8CM) BRAIDED INJ LINE, ROTATING ADAPT, List Number MX682BRJanuary 24, 2024
Questions about this recall
Was HIGH PRESSURE ROTATOR WITH MALE LUER LOCK, List Number MX496HP recalled?
Yes. FDA published recall Z-0743-2024 on January 24, 2024. Reported hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
What should I do if I have this product?
Voluntary: Firm initiated · Letter