HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when connected to the HeartMate 3 Left Ventricular Assist System, REF 106524US, UDI or GTIN: 00813024013297, and 106524, UDI or GTIN: 00813024011712 - Product Usage: Provides power to the System Controller and pump. - Provides power to the System Monitor when it is connected to the Power Module. - Connects the System Monitor to the System Controller for monitoring purposes. - Echoes System Controller alarms. The Mobile Power Unit is for home or clinical use when the patient does not require monitoring using the System Monitor. The Mobile Power Unit is used when the patient is indoors, stationary, or sleeping. The System Controller and the Mobile Power Unit are connected through the Mobile Power Unit patient cable. The cable transfers power from the Mobile Power Unit to the System Controller.
Published January 8, 2020 · Recall initiated: December 2, 2019
Summary
FDA records medical device Class II recall Z-0746-2020, published January 8, 2020 and initiated December 2, 2019. Recalling firm: Thoratec Corp.; Status: Terminated; Reported quantity: 13871; Lot: All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Excessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which could lead to no or reduced blood flow from the left ventricular assist system.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Thoratec Corp.
- Distributor
- US Nationwide distribution in the states of MN, PA, IL, NJ, WI, AZ, AR, TN, MO, TX, MA, NE, MI, CA, NC, CO, OH, DE, FL, NY, IN, GA, VA, DC, CT, OK, UT, KY, MD, OR, NM, ME, SC, IA, WA, LA, NV, AL, KS, MS, WV.
- UPC
- 00813024010883
- Lot
- All serial numbers.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386,June 19, 2024
- FDAThoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnlyMay 8, 2024
- FDAThoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnlyMay 8, 2024
- FDAHeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief,March 27, 2024
- FDAGreat Value Macadamia Nuts 4 oz. pouch with UPC CODE: 078742309422 packaged in a cello bag, 6 per caseJanuary 8, 2020
- FDAMedicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM xJanuary 8, 2020
Questions about this recall
Was HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when connected to the HeartMate 3 Left Ventricular Assist System, REF 106524US, UDI or GTIN: 00813024013297, and 106524, UDI or GTIN: 00813024011712 - Product Usage: Provides power to the System Controller and pump. - Provides power to the System Monitor when it is connected to the Power Module. - Connects the System Monitor to the System Controller for monitoring purposes. - Echoes System Controller alarms. The Mobile Power Unit is for home or clinical use when the patient does not require monitoring using the System Monitor. The Mobile Power Unit is used when the patient is indoors, stationary, or sleeping. The System Controller and the Mobile Power Unit are connected through the Mobile Power Unit patient cable. The cable transfers power from the Mobile Power Unit to the System Controller. recalled?
Yes. FDA published recall Z-0746-2020 on January 8, 2020. Reported hazard: Excessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which could lead to no or reduced blood flow from the left ventricular assist system.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 00813024010883.