Back to search
FDAFDA · Food and drugs·Recall number Z-0746-2024

Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX682R

Published January 24, 2024 · Recall initiated: November 10, 2023

Summary

FDA records medical device Class II recall Z-0746-2024, published January 24, 2024 and initiated November 10, 2023. Recalling firm: Smiths Medical Asd Inc; Status: Ongoing; Reported quantity: 31,685 total; Lot: Lot Number: 4398433. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical Asd Inc
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, UAE.
Lot
Lot Number: 4398433
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX682R recalled?

Yes. FDA published recall Z-0746-2024 on January 24, 2024. Reported hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

What should I do if I have this product?

Voluntary: Firm initiated · Letter