Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX682R
Published January 24, 2024 · Recall initiated: November 10, 2023
Summary
FDA records medical device Class II recall Z-0746-2024, published January 24, 2024 and initiated November 10, 2023. Recalling firm: Smiths Medical Asd Inc; Status: Ongoing; Reported quantity: 31,685 total; Lot: Lot Number: 4398433. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Smiths Medical Asd Inc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, UAE.
- Lot
- Lot Number: 4398433
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694RJanuary 24, 2024
- FDAHIGH PRESSURE ROTATOR WITH MALE LUER LOCK, List Number MX496HPJanuary 24, 2024
- FDAHIGH PRESSURE ROTATOR WITH FEMALE LUER LOCK, List Number MX497HPJanuary 24, 2024
- FDA6IN PRESSURE TUBING, List Number MX20617January 24, 2024
- FDA20IN (50.8CM) BRAIDED INJ LINE, ROTATING ADAPT, List Number MX682BRJanuary 24, 2024
- FDASOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S,January 24, 2024
Questions about this recall
Was Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX682R recalled?
Yes. FDA published recall Z-0746-2024 on January 24, 2024. Reported hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
What should I do if I have this product?
Voluntary: Firm initiated · Letter