GE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants.

Published February 20, 2013 · Recall initiated: October 15, 2012

Summary

FDA records medical device Class I recall Z-0748-2013, published February 20, 2013 and initiated October 15, 2012. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 276 units total, 109 of those nationwide.; Lot: Models M1139417, M1223628, M1192226-072951, M1192226-084200. M1192226-095181, M1192226-104599. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Worldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA, MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX, VT, WA, WI and WV; and the countries of Australia, Belgium, Canada, China, France, New Zealand, Poland, Saudi Arabia, Switzerland and United Kingdom.
Lot
Models M1139417, M1223628, M1192226-072951, M1192226-084200. M1192226-095181, M1192226-104599
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. recalled?

Yes. FDA published recall Z-0748-2013 on February 20, 2013. Reported hazard: Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

What should I do if I have this product?

Voluntary: Firm initiated · Letter