GE Healthcare, Optima XR220amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.
Published February 17, 2016 · Recall initiated: December 23, 2015
Summary
FDA records medical device Class II recall Z-0768-2016, published February 17, 2016 and initiated December 23, 2015. Recalling firm: GE Medical Systems, LLC; Status: Terminated; Reported quantity: 2389; Lot: Lot or Serial numbers including but not limited to: 00001033715WK2, 12535HL3, 00000078321HL7, 00000096815HL6, 00000081147HL1, 00001032621WK3, 00000092076HL9, 00000092382HL1, 00000460115GE9, 00000079932HL0, D3s14793, 00000088642HL4, 116635HL4, 120632HL5, GEGR2411, GEGR5073, 00001027151WK8, 00000129105HL3, 261417HP0, 261416HP2, 1035189WK8, 00000084720HL2, 00000087109HL5, 00000100735HL0, 121206HL7, 125007HL5, 125019HL0, 1028231WK7, 00000126237HL7, 00000102072HL6, 00000092394HL6, 00000115802HL1, 124167HL8, 00001033721WK0, 00000108520HL8, 00000087928HL8, 00000111794HL4, 00000110657HL4, 00000092629HL5, 119090HL9, 00000442129GE3, 00000128179HL9, 00000110710HL1, 83621HL3, 00000082207HL2, 1034762WK3, 00000081499HL6, 00000088705HL9, 00000085601HL3, 00001031247WK8, 00000084721HL0, 00000084410HL0, 00000086179HL9, 1033866WK3, 00000084286HL4, 00000096817HL2, 00000092948HL9, 00000106841HL0, 00000105475HL8, 00001033220WK3, 00000106072HL2, 00000110662HL4, 00000106055HL7, 00000008789HL5, 00000084415HL9, 00000084713HL7, 00000091763HL3, 00000093367HL1, 123730HL4, 00000109436HL6, 00000078816HL6, 00000078490HL0, 120816HL4, 00000124501HL8, 00000124510HL9, 123456789, 00000095656BT5, 118267HL4, 00000081688HL4, 00000110668HL1, 123737HL9, 00000081500HL1, 00000084711HL1, 00000085596HL5, 00000118284HL9, 00000124201HL5, 00000118285HL6, 00000118462HL1, 00000118282HL3, 00000104908HL9, 00000102090HL8, 00000100174HL2, 120182HL1, D3S141084, 00000100430HL8, 00000097754HL6, 114137HL3, 00000099201HL6, 00000099204HL0, 00000111427HL1, 00000087073HL3, 120184HL7, 00000126701HL2, 00000128172HL4, 00000124478HL9, 00000124224HL7, 00000091044HL8, 00000107457HL4, 00000076792HL1, 00000112148HL2, 00001032707WK0, 00000101689HL8, 119104HL8, 00001027765WK5, 114191HL0, 00000085591HL6, 00001030627WK2, 00001030293WK3, 00001030654WK6, 00001030617WK3, GEGR3510, 119099HL0, 00000106829HL5, 120187HL0, 00000087076HL6, 00000124186HL8, 00000098182HL9, 120829HL7, 00000093383HL8, 124217HL1, 122883HL2, 121230HL7, 112145HL8, 00001033608WK9, 00000096582HL2, 121229hl9, 00000081796HL5, 00000404458GE2, 00000076912HL5, 00000080863HL4, 00000108423HL5, 00001030034WK1, 120159HL9, 00001029071WK6, 00000093374HL7, 00000087107HL9, 00001031749WK3, 00000091761HL7, 00000086996HL6, 00000092073HL6, 00001035466WK0, 133379BT8, 00001034661WK7, GEGR2475, 00000081817HL9, 00000124181HL9, 00000111919HL7, 00000109407HL7, 00000105795HL9, 00000106056HL5, 00000116630HL5, 00001027347WK2, 00000083537HL1, 00000086155HL9, 00000078491HL8, GEGR3252, 00000089859HL3, 00000096580HL6, 15B1433, 00000077251HL7, 00001031962WK2, 00001035186WK4, 00000081496HL2, 00000088645HL7, 00000080783HL4, 120808HL1, 00000087094HL9, 00000105455HL0, 00000119756HL5, 00001035456WK1, 00000083015HL8, 00000082740HL2, 00000084147HL8, 4123, 00000085042HL0, 00000087890HL0, 00001029042WK7, 00001031728WK7, 00000076403HL5, 00000092392HL0, 00000081819HL5, 00000125677HL5, 00000149748BT6, 00000101296HL2, 00000109423HL4, 00000108571HL1, 00000109425HL9, 00000090229HL6, 00000089140HL8, 00000085904HL1, 00000087922HL1, T600342, 00000093400HL0, 117199HL0, 00001033662WK6, 5331287-2, 113151hl5, 00000091758HL3, 00000095898HL3, UA46793-8, 00000100423HL3, 00000098211HL6, 1027422WK3, 116643HL8, 103908WK3, 00000110649HL1, 00000087077HL4, 00000084168HL4, 00000108418HL5, 108417hl7, 142904BT2, 116656HL0, 00000101693HL0, 00000122245HL4, 00001032825WX0, 1032825WK0, 00000081684HL3, 120678HL8, 00000078815HL8, 00000080777HL6, 00000080859HL2, 00000082462HL3, 00000083541HL3, 00000082725HL3, 00000082744HL4, 00000080633HL1, 00000085053HL7, 00000107167HL9, 00000092946HL3, 00000088702HL6, 00000089493HL1, 00000104904HL8, 118690HL7, 118689HL9, 122947HL5, 120655HL6, 121199HL4, 122031HL8, 120834HL7, 00000458240GE9, 124189HL2, 00000106047HL4, 00000120195HL3, 00000114150HL6, 124009HL2, 00000085913HL2, 119091hl7, 00000129120HL2, 00000129116HL0, 00000102997HL4, 00000101970HL2, 00000106076HL3, 00000079924HL7, 00000086169HL0, 00000124182HL7, 00000105457HL6, 00000098190HL2, 00000081813HL8, 0000008301. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Medical Systems, LLC
- Distributor
- Worldwide Distribution -- US, including Washington D.C. Puerto Rico, Virgin Islands, and Guam; and the countries of Afghanistan, Albania, Anguilla, Antigua and Barbuda, Armenia, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, Ethiopia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Lebanon, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Papua New Guinea, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Trinidad , Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
- Lot
- Lot or Serial numbers including but not limited to: 00001033715WK2, 12535HL3, 00000078321HL7, 00000096815HL6, 00000081147HL1, 00001032621WK3, 00000092076HL9, 00000092382HL1, 00000460115GE9, 00000079932HL0, D3s14793, 00000088642HL4, 116635HL4, 120632HL5, GEGR2411, GEGR5073, 00001027151WK8, 00000129105HL3, 261417HP0, 261416HP2, 1035189WK8, 00000084720HL2, 00000087109HL5, 00000100735HL0, 121206HL7, 125007HL5, 125019HL0, 1028231WK7, 00000126237HL7, 00000102072HL6, 00000092394HL6, 00000115802HL1, 124167HL8, 00001033721WK0, 00000108520HL8, 00000087928HL8, 00000111794HL4, 00000110657HL4, 00000092629HL5, 119090HL9, 00000442129GE3, 00000128179HL9, 00000110710HL1, 83621HL3, 00000082207HL2, 1034762WK3, 00000081499HL6, 00000088705HL9, 00000085601HL3, 00001031247WK8, 00000084721HL0, 00000084410HL0, 00000086179HL9, 1033866WK3, 00000084286HL4, 00000096817HL2, 00000092948HL9, 00000106841HL0, 00000105475HL8, 00001033220WK3, 00000106072HL2, 00000110662HL4, 00000106055HL7, 00000008789HL5, 00000084415HL9, 00000084713HL7, 00000091763HL3, 00000093367HL1, 123730HL4, 00000109436HL6, 00000078816HL6, 00000078490HL0, 120816HL4, 00000124501HL8, 00000124510HL9, 123456789, 00000095656BT5, 118267HL4, 00000081688HL4, 00000110668HL1, 123737HL9, 00000081500HL1, 00000084711HL1, 00000085596HL5, 00000118284HL9, 00000124201HL5, 00000118285HL6, 00000118462HL1, 00000118282HL3, 00000104908HL9, 00000102090HL8, 00000100174HL2, 120182HL1, D3S141084, 00000100430HL8, 00000097754HL6, 114137HL3, 00000099201HL6, 00000099204HL0, 00000111427HL1, 00000087073HL3, 120184HL7, 00000126701HL2, 00000128172HL4, 00000124478HL9, 00000124224HL7, 00000091044HL8, 00000107457HL4, 00000076792HL1, 00000112148HL2, 00001032707WK0, 00000101689HL8, 119104HL8, 00001027765WK5, 114191HL0, 00000085591HL6, 00001030627WK2, 00001030293WK3, 00001030654WK6, 00001030617WK3, GEGR3510, 119099HL0, 00000106829HL5, 120187HL0, 00000087076HL6, 00000124186HL8, 00000098182HL9, 120829HL7, 00000093383HL8, 124217HL1, 122883HL2, 121230HL7, 112145HL8, 00001033608WK9, 00000096582HL2, 121229hl9, 00000081796HL5, 00000404458GE2, 00000076912HL5, 00000080863HL4, 00000108423HL5, 00001030034WK1, 120159HL9, 00001029071WK6, 00000093374HL7, 00000087107HL9, 00001031749WK3, 00000091761HL7, 00000086996HL6, 00000092073HL6, 00001035466WK0, 133379BT8, 00001034661WK7, GEGR2475, 00000081817HL9, 00000124181HL9, 00000111919HL7, 00000109407HL7, 00000105795HL9, 00000106056HL5, 00000116630HL5, 00001027347WK2, 00000083537HL1, 00000086155HL9, 00000078491HL8, GEGR3252, 00000089859HL3, 00000096580HL6, 15B1433, 00000077251HL7, 00001031962WK2, 00001035186WK4, 00000081496HL2, 00000088645HL7, 00000080783HL4, 120808HL1, 00000087094HL9, 00000105455HL0, 00000119756HL5, 00001035456WK1, 00000083015HL8, 00000082740HL2, 00000084147HL8, 4123, 00000085042HL0, 00000087890HL0, 00001029042WK7, 00001031728WK7, 00000076403HL5, 00000092392HL0, 00000081819HL5, 00000125677HL5, 00000149748BT6, 00000101296HL2, 00000109423HL4, 00000108571HL1, 00000109425HL9, 00000090229HL6, 00000089140HL8, 00000085904HL1, 00000087922HL1, T600342, 00000093400HL0, 117199HL0, 00001033662WK6, 5331287-2, 113151hl5, 00000091758HL3, 00000095898HL3, UA46793-8, 00000100423HL3, 00000098211HL6, 1027422WK3, 116643HL8, 103908WK3, 00000110649HL1, 00000087077HL4, 00000084168HL4, 00000108418HL5, 108417hl7, 142904BT2, 116656HL0, 00000101693HL0, 00000122245HL4, 00001032825WX0, 1032825WK0, 00000081684HL3, 120678HL8, 00000078815HL8, 00000080777HL6, 00000080859HL2, 00000082462HL3, 00000083541HL3, 00000082725HL3, 00000082744HL4, 00000080633HL1, 00000085053HL7, 00000107167HL9, 00000092946HL3, 00000088702HL6, 00000089493HL1, 00000104904HL8, 118690HL7, 118689HL9, 122947HL5, 120655HL6, 121199HL4, 122031HL8, 120834HL7, 00000458240GE9, 124189HL2, 00000106047HL4, 00000120195HL3, 00000114150HL6, 124009HL2, 00000085913HL2, 119091hl7, 00000129120HL2, 00000129116HL0, 00000102997HL4, 00000101970HL2, 00000106076HL3, 00000079924HL7, 00000086169HL0, 00000124182HL7, 00000105457HL6, 00000098190HL2, 00000081813HL8, 0000008301
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASenographe Pristina 3DNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDAOptima XR240amx, Mobile Digital Radiographic SystemJuly 22, 2026
- FDAMEVION S250 Radiation therapy systemFebruary 17, 2016
- FDAIMPLANT,TM, MTX 4.1 mm x 11.5 mm diameter 3.5 mm Catalog #TMM4B11 Product Usage: Zimmer Dental implant systemsFebruary 17, 2016
Questions about this recall
Was GE Healthcare, Optima XR220amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images. recalled?
Yes. FDA published recall Z-0768-2016 on February 17, 2016. Reported hazard: Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
What should I do if I have this product?
Voluntary: Firm initiated · Letter