GE Healthcare, Optima XR220amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.

Published February 17, 2016 · Recall initiated: December 23, 2015

Summary

FDA records medical device Class II recall Z-0768-2016, published February 17, 2016 and initiated December 23, 2015. Recalling firm: GE Medical Systems, LLC; Status: Terminated; Reported quantity: 2389; Lot: Lot or Serial numbers including but not limited to: 00001033715WK2, 12535HL3, 00000078321HL7, 00000096815HL6, 00000081147HL1, 00001032621WK3, 00000092076HL9, 00000092382HL1, 00000460115GE9, 00000079932HL0, D3s14793, 00000088642HL4, 116635HL4, 120632HL5, GEGR2411, GEGR5073, 00001027151WK8, 00000129105HL3, 261417HP0, 261416HP2, 1035189WK8, 00000084720HL2, 00000087109HL5, 00000100735HL0, 121206HL7, 125007HL5, 125019HL0, 1028231WK7, 00000126237HL7, 00000102072HL6, 00000092394HL6, 00000115802HL1, 124167HL8, 00001033721WK0, 00000108520HL8, 00000087928HL8, 00000111794HL4, 00000110657HL4, 00000092629HL5, 119090HL9, 00000442129GE3, 00000128179HL9, 00000110710HL1, 83621HL3, 00000082207HL2, 1034762WK3, 00000081499HL6, 00000088705HL9, 00000085601HL3, 00001031247WK8, 00000084721HL0, 00000084410HL0, 00000086179HL9, 1033866WK3, 00000084286HL4, 00000096817HL2, 00000092948HL9, 00000106841HL0, 00000105475HL8, 00001033220WK3, 00000106072HL2, 00000110662HL4, 00000106055HL7, 00000008789HL5, 00000084415HL9, 00000084713HL7, 00000091763HL3, 00000093367HL1, 123730HL4, 00000109436HL6, 00000078816HL6, 00000078490HL0, 120816HL4, 00000124501HL8, 00000124510HL9, 123456789, 00000095656BT5, 118267HL4, 00000081688HL4, 00000110668HL1, 123737HL9, 00000081500HL1, 00000084711HL1, 00000085596HL5, 00000118284HL9, 00000124201HL5, 00000118285HL6, 00000118462HL1, 00000118282HL3, 00000104908HL9, 00000102090HL8, 00000100174HL2, 120182HL1, D3S141084, 00000100430HL8, 00000097754HL6, 114137HL3, 00000099201HL6, 00000099204HL0, 00000111427HL1, 00000087073HL3, 120184HL7, 00000126701HL2, 00000128172HL4, 00000124478HL9, 00000124224HL7, 00000091044HL8, 00000107457HL4, 00000076792HL1, 00000112148HL2, 00001032707WK0, 00000101689HL8, 119104HL8, 00001027765WK5, 114191HL0, 00000085591HL6, 00001030627WK2, 00001030293WK3, 00001030654WK6, 00001030617WK3, GEGR3510, 119099HL0, 00000106829HL5, 120187HL0, 00000087076HL6, 00000124186HL8, 00000098182HL9, 120829HL7, 00000093383HL8, 124217HL1, 122883HL2, 121230HL7, 112145HL8, 00001033608WK9, 00000096582HL2, 121229hl9, 00000081796HL5, 00000404458GE2, 00000076912HL5, 00000080863HL4, 00000108423HL5, 00001030034WK1, 120159HL9, 00001029071WK6, 00000093374HL7, 00000087107HL9, 00001031749WK3, 00000091761HL7, 00000086996HL6, 00000092073HL6, 00001035466WK0, 133379BT8, 00001034661WK7, GEGR2475, 00000081817HL9, 00000124181HL9, 00000111919HL7, 00000109407HL7, 00000105795HL9, 00000106056HL5, 00000116630HL5, 00001027347WK2, 00000083537HL1, 00000086155HL9, 00000078491HL8, GEGR3252, 00000089859HL3, 00000096580HL6, 15B1433, 00000077251HL7, 00001031962WK2, 00001035186WK4, 00000081496HL2, 00000088645HL7, 00000080783HL4, 120808HL1, 00000087094HL9, 00000105455HL0, 00000119756HL5, 00001035456WK1, 00000083015HL8, 00000082740HL2, 00000084147HL8, 4123, 00000085042HL0, 00000087890HL0, 00001029042WK7, 00001031728WK7, 00000076403HL5, 00000092392HL0, 00000081819HL5, 00000125677HL5, 00000149748BT6, 00000101296HL2, 00000109423HL4, 00000108571HL1, 00000109425HL9, 00000090229HL6, 00000089140HL8, 00000085904HL1, 00000087922HL1, T600342, 00000093400HL0, 117199HL0, 00001033662WK6, 5331287-2, 113151hl5, 00000091758HL3, 00000095898HL3, UA46793-8, 00000100423HL3, 00000098211HL6, 1027422WK3, 116643HL8, 103908WK3, 00000110649HL1, 00000087077HL4, 00000084168HL4, 00000108418HL5, 108417hl7, 142904BT2, 116656HL0, 00000101693HL0, 00000122245HL4, 00001032825WX0, 1032825WK0, 00000081684HL3, 120678HL8, 00000078815HL8, 00000080777HL6, 00000080859HL2, 00000082462HL3, 00000083541HL3, 00000082725HL3, 00000082744HL4, 00000080633HL1, 00000085053HL7, 00000107167HL9, 00000092946HL3, 00000088702HL6, 00000089493HL1, 00000104904HL8, 118690HL7, 118689HL9, 122947HL5, 120655HL6, 121199HL4, 122031HL8, 120834HL7, 00000458240GE9, 124189HL2, 00000106047HL4, 00000120195HL3, 00000114150HL6, 124009HL2, 00000085913HL2, 119091hl7, 00000129120HL2, 00000129116HL0, 00000102997HL4, 00000101970HL2, 00000106076HL3, 00000079924HL7, 00000086169HL0, 00000124182HL7, 00000105457HL6, 00000098190HL2, 00000081813HL8, 0000008301. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Medical Systems, LLC
Distributor
Worldwide Distribution -- US, including Washington D.C. Puerto Rico, Virgin Islands, and Guam; and the countries of Afghanistan, Albania, Anguilla, Antigua and Barbuda, Armenia, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, Ethiopia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Lebanon, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Papua New Guinea, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Trinidad , Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Lot
Lot or Serial numbers including but not limited to: 00001033715WK2, 12535HL3, 00000078321HL7, 00000096815HL6, 00000081147HL1, 00001032621WK3, 00000092076HL9, 00000092382HL1, 00000460115GE9, 00000079932HL0, D3s14793, 00000088642HL4, 116635HL4, 120632HL5, GEGR2411, GEGR5073, 00001027151WK8, 00000129105HL3, 261417HP0, 261416HP2, 1035189WK8, 00000084720HL2, 00000087109HL5, 00000100735HL0, 121206HL7, 125007HL5, 125019HL0, 1028231WK7, 00000126237HL7, 00000102072HL6, 00000092394HL6, 00000115802HL1, 124167HL8, 00001033721WK0, 00000108520HL8, 00000087928HL8, 00000111794HL4, 00000110657HL4, 00000092629HL5, 119090HL9, 00000442129GE3, 00000128179HL9, 00000110710HL1, 83621HL3, 00000082207HL2, 1034762WK3, 00000081499HL6, 00000088705HL9, 00000085601HL3, 00001031247WK8, 00000084721HL0, 00000084410HL0, 00000086179HL9, 1033866WK3, 00000084286HL4, 00000096817HL2, 00000092948HL9, 00000106841HL0, 00000105475HL8, 00001033220WK3, 00000106072HL2, 00000110662HL4, 00000106055HL7, 00000008789HL5, 00000084415HL9, 00000084713HL7, 00000091763HL3, 00000093367HL1, 123730HL4, 00000109436HL6, 00000078816HL6, 00000078490HL0, 120816HL4, 00000124501HL8, 00000124510HL9, 123456789, 00000095656BT5, 118267HL4, 00000081688HL4, 00000110668HL1, 123737HL9, 00000081500HL1, 00000084711HL1, 00000085596HL5, 00000118284HL9, 00000124201HL5, 00000118285HL6, 00000118462HL1, 00000118282HL3, 00000104908HL9, 00000102090HL8, 00000100174HL2, 120182HL1, D3S141084, 00000100430HL8, 00000097754HL6, 114137HL3, 00000099201HL6, 00000099204HL0, 00000111427HL1, 00000087073HL3, 120184HL7, 00000126701HL2, 00000128172HL4, 00000124478HL9, 00000124224HL7, 00000091044HL8, 00000107457HL4, 00000076792HL1, 00000112148HL2, 00001032707WK0, 00000101689HL8, 119104HL8, 00001027765WK5, 114191HL0, 00000085591HL6, 00001030627WK2, 00001030293WK3, 00001030654WK6, 00001030617WK3, GEGR3510, 119099HL0, 00000106829HL5, 120187HL0, 00000087076HL6, 00000124186HL8, 00000098182HL9, 120829HL7, 00000093383HL8, 124217HL1, 122883HL2, 121230HL7, 112145HL8, 00001033608WK9, 00000096582HL2, 121229hl9, 00000081796HL5, 00000404458GE2, 00000076912HL5, 00000080863HL4, 00000108423HL5, 00001030034WK1, 120159HL9, 00001029071WK6, 00000093374HL7, 00000087107HL9, 00001031749WK3, 00000091761HL7, 00000086996HL6, 00000092073HL6, 00001035466WK0, 133379BT8, 00001034661WK7, GEGR2475, 00000081817HL9, 00000124181HL9, 00000111919HL7, 00000109407HL7, 00000105795HL9, 00000106056HL5, 00000116630HL5, 00001027347WK2, 00000083537HL1, 00000086155HL9, 00000078491HL8, GEGR3252, 00000089859HL3, 00000096580HL6, 15B1433, 00000077251HL7, 00001031962WK2, 00001035186WK4, 00000081496HL2, 00000088645HL7, 00000080783HL4, 120808HL1, 00000087094HL9, 00000105455HL0, 00000119756HL5, 00001035456WK1, 00000083015HL8, 00000082740HL2, 00000084147HL8, 4123, 00000085042HL0, 00000087890HL0, 00001029042WK7, 00001031728WK7, 00000076403HL5, 00000092392HL0, 00000081819HL5, 00000125677HL5, 00000149748BT6, 00000101296HL2, 00000109423HL4, 00000108571HL1, 00000109425HL9, 00000090229HL6, 00000089140HL8, 00000085904HL1, 00000087922HL1, T600342, 00000093400HL0, 117199HL0, 00001033662WK6, 5331287-2, 113151hl5, 00000091758HL3, 00000095898HL3, UA46793-8, 00000100423HL3, 00000098211HL6, 1027422WK3, 116643HL8, 103908WK3, 00000110649HL1, 00000087077HL4, 00000084168HL4, 00000108418HL5, 108417hl7, 142904BT2, 116656HL0, 00000101693HL0, 00000122245HL4, 00001032825WX0, 1032825WK0, 00000081684HL3, 120678HL8, 00000078815HL8, 00000080777HL6, 00000080859HL2, 00000082462HL3, 00000083541HL3, 00000082725HL3, 00000082744HL4, 00000080633HL1, 00000085053HL7, 00000107167HL9, 00000092946HL3, 00000088702HL6, 00000089493HL1, 00000104904HL8, 118690HL7, 118689HL9, 122947HL5, 120655HL6, 121199HL4, 122031HL8, 120834HL7, 00000458240GE9, 124189HL2, 00000106047HL4, 00000120195HL3, 00000114150HL6, 124009HL2, 00000085913HL2, 119091hl7, 00000129120HL2, 00000129116HL0, 00000102997HL4, 00000101970HL2, 00000106076HL3, 00000079924HL7, 00000086169HL0, 00000124182HL7, 00000105457HL6, 00000098190HL2, 00000081813HL8, 0000008301
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GE Healthcare, Optima XR220amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images. recalled?

Yes. FDA published recall Z-0768-2016 on February 17, 2016. Reported hazard: Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.

What should I do if I have this product?

Voluntary: Firm initiated · Letter