ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM SS/SS ECHO B PG, Item Number H787065970395, Catalog No.REF 06597039, 2) 5 F Stiffened Micro-Introducer Kit, STF 5F M.I. KIT 45CM SS/SS ECHO S PG, Item Number H787065970415, Catalog No. REF 06597041 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick
Published December 28, 2016 · Recall initiated: September 28, 2016
Summary
FDA records medical device Class III recall Z-0772-2017, published December 28, 2016 and initiated September 28, 2016. Recalling firm: Angiodynamics, Inc.; Status: Terminated; Reported quantity: 162 boxes; Lot: Batch/Lot: 1) 5046426, 2) 5022317, 5032458, 5040672, 5046428, 5063983 & 5081983. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Angiodynamics, Inc.
- Distributor
- Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
- Lot
- Batch/Lot: 1) 5046426, 2) 5022317, 5032458, 5040672, 5046428, 5063983 & 5081983
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAccuVu Angiographic Catheter, Straight Flush, Non-Braided, With 10 R/O Markers; ACCU-VU STR FL 4F X 100CM 035April 29, 2026
- FDASoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 65CM 038 NB 0SH; Catalog No.: 10714014;April 29, 2026
- FDASOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701August 7, 2024
- FDASOFT-VU KUMPE 5F X 40CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732702August 7, 2024
- FDAStreit's Quinoa with Mushrooms & Herbs, Net Wt. 7.7 oz (220g). Product of IsraelDecember 28, 2016
- FDAANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit,December 28, 2016
Questions about this recall
Was ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM SS/SS ECHO B PG, Item Number H787065970395, Catalog No.REF 06597039, 2) 5 F Stiffened Micro-Introducer Kit, STF 5F M.I. KIT 45CM SS/SS ECHO S PG, Item Number H787065970415, Catalog No. REF 06597041 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick recalled?
Yes. FDA published recall Z-0772-2017 on December 28, 2016. Reported hazard: Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.
What should I do if I have this product?
Voluntary: Firm initiated · Letter