Back to search
FDAFDA · Food and drugs·Recall number Z-0777-2017

ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T NON-ECHO B PG, Item Number H787065970685, Catalog No.REF 06597068, 2) 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T NON-ECHO S PG, Item Number H787065970705, Catalog No.REF 06597070 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick

Published December 28, 2016 · Recall initiated: September 28, 2016

Summary

FDA records medical device Class II recall Z-0777-2017, published December 28, 2016 and initiated September 28, 2016. Recalling firm: Angiodynamics, Inc.; Status: Terminated; Reported quantity: 7 boxes; Lot: Batch/Lot: 1) 5034280, 5067423, 2) 5061461, 5070972. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Angiodynamics, Inc.
Distributor
Nationwide Distribution to AL, AR, CA, FL, GA, IL, KY, LA, MI, MN, MO, NC, OH, OK, SC, NY, ME, PA, MD, VA, WA, WI, WV, TN, and TX
Lot
Batch/Lot: 1) 5034280, 5067423, 2) 5061461, 5070972
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T NON-ECHO B PG, Item Number H787065970685, Catalog No.REF 06597068, 2) 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T NON-ECHO S PG, Item Number H787065970705, Catalog No.REF 06597070 Indicated for percutaneous introduction of guidewire or catheter into the vascular system following a small 21-gauge needle stick recalled?

Yes. FDA published recall Z-0777-2017 on December 28, 2016. Reported hazard: Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

What should I do if I have this product?

Voluntary: Firm initiated · Letter