Back to search
FDAFDA · Food and drugs·Recall number Z-0778-2016

IMPLANT,TM, MTX FULL 4.1 mm x 10 mm diameter 3.5 mm Catalog #TMT4B10 Product Usage: Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.

Published February 17, 2016 · Recall initiated: January 6, 2016

Summary

FDA records medical device Class II recall Z-0778-2016, published February 17, 2016 and initiated January 6, 2016. Recalling firm: Zimmer Dental Inc; Status: Terminated; Reported quantity: 1059; Lot: Lot # 63187148 63192992 63212188 63216161. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer Dental Inc
Distributor
Worldwide Distribution. US Nationwide in the states of NY, CA, KS, NJ, CO, MI, MD, WA, CT, KY, OR, OH, IA, SC, AZ, AK, IN, TN, WI, FL, AL, UT, WY, MA, PA, MN, IL, MI, GA, ND, RI, including Puerto Rico; and the countries of Thailand, France, Germany, Colombia, Hungary, Netherlands, Saudi Arabia, Poland, Spain, and Italy
Lot
Lot # 63187148 63192992 63212188 63216161
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was IMPLANT,TM, MTX FULL 4.1 mm x 10 mm diameter 3.5 mm Catalog #TMT4B10 Product Usage: Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. recalled?

Yes. FDA published recall Z-0778-2016 on February 17, 2016. Reported hazard: Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.

What should I do if I have this product?

Voluntary: Firm initiated · Letter