FDAFDA · Food and drugs·Recall number Z-0780-2014
XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
Published January 29, 2014 · Recall initiated: November 25, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Elekta, Inc.
- Distributor
- Worldwide Distribution - US Distribution including the states of IL, IN, LA, NJ, TX, WA and WI., and the countries of Belgium, Germany, Italy and Netherlands.
- Lot
- Model or Catalog No. MUJ; Software Version XiO Versions 4.80 and higher
- Risk classification
- Class II