FDAFDA · Food and drugs·Recall number Z-0785-2014

ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

Published January 29, 2014 · Recall initiated: December 4, 2013

Summary

FDA records medical device Class II recall Z-0785-2014, published January 29, 2014 and initiated December 4, 2013. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 255 units; Lot: ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005, 9510027, 9510028, 9603042, 9603043, 9604044, 9604047, 9605011, 9607060, 9609065, 9610072, 9704081, 9709090, 9710097, 9711103, 9712105, 9712106, 9801107, 9801109, 9802112, 9803001, 9803116, 9804126, 9804129, 9806136, 9807138, 9807422,9810527, 9901156, 9901160, 9901162, 9908167, 9909172, 9910176, 00709B, 0101003A, 01310B, 01525B, 01920B, 01921B, 01922B, 03041B, 03094B, 03338B, 03368B, 2145-30074A, 7718B, A0103015, A0103213, A0105217, A02100230, A02100231, A02110248, A02110249, A02110250, A02110251, A02120254, A02120256, A03010262, A03010264, A03010265, A03010266, A03010270, A03020274, A03020275, A03030282, A03070283, A03120294, B0003196, B0007200, B0309014, ID3935B, M@10639, M@10645, M@10774, M@10777, M@10877, M@11239, M@11568, M@11640, M@11684, M@11685, M@11729, M@11769, M@11813, M@11947, M@12378-1, M@12462, M@12765, M@13725, M@24470, M@4750028851, M@4750028853, M@5348B, M@5807B, M@5874B, M@5894B, M@5932B, M@5941B, M@5963B, M@6041B, M@6062B, M@6229B, M@6354B, M@6477B, M@6485B, M@6546B, M@6555B, M@6692B, M@6754B, M@6809B, M@6815B, M@6948B, M@7007B, M@7023B, M@7093B, M@7194B, M@7198B, M@7207B, M@7214B, M@7280B, M@7285B, M@7286B, M@7408B, M@7540B, M@7596B, M@7675B, M@7704B, M@8156B, M@8324B, M@8344B, M@8429B, M@8523B, M@8578B, M@8588B, M@8637B, M@8639RB, M@8648B, M@8808B, M@8813B, M@8898B, M@8977B, M@9050B, M@9261B, M@9276B, M@9355RB, M@9475B, M@9544B, M@9647B, M@9656B, M@9733B & M@9796B. ADAC ARGUS: Model #88295; Serial #: 9211014, 9301005, 9304019, 9304024, 9304028, 9401028, 9401031, 9401037, 9505014, 9910033, 93-10678- 1292, 9910028P, M@10225AMT, M@3732B, M@3903B, M@3954B, M@3956B, M@3960B, M@4249B, M@4415B, M@4646B, M@4868B, M@5172B, M@5321B, M@7341B & M@AMT1154. ADAC CIRRUS: Model #88294; 34, 718558, 9206009, 9207003, 9207004, 9207005, 9207007, 9211006, 9300113, 9304013, 9309012, 9404022, 9407028, 9509042, 9604098, 9910030, 93-10735-0393, 9621RQ, M@10110AMT, M@10324AMT, M@10402AMT, M@3599G-08, M@3905Q, M@4029Q, M@4129Q, M@4142Q, M@4668Q, M@4743Q, M@5003Q, M@5668Q, M@5895Q, M@AMT126R, M@AMT1282 & M@MVS3227A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution: US (nationwide) including states of: AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great Britain, Italy, Netherlands, Norway, Pakistan, Russia, Singapore & South Korea.
Lot
ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005, 9510027, 9510028, 9603042, 9603043, 9604044, 9604047, 9605011, 9607060, 9609065, 9610072, 9704081, 9709090, 9710097, 9711103, 9712105, 9712106, 9801107, 9801109, 9802112, 9803001, 9803116, 9804126, 9804129, 9806136, 9807138, 9807422,9810527, 9901156, 9901160, 9901162, 9908167, 9909172, 9910176, 00709B, 0101003A, 01310B, 01525B, 01920B, 01921B, 01922B, 03041B, 03094B, 03338B, 03368B, 2145-30074A, 7718B, A0103015, A0103213, A0105217, A02100230, A02100231, A02110248, A02110249, A02110250, A02110251, A02120254, A02120256, A03010262, A03010264, A03010265, A03010266, A03010270, A03020274, A03020275, A03030282, A03070283, A03120294, B0003196, B0007200, B0309014, ID3935B, M@10639, M@10645, M@10774, M@10777, M@10877, M@11239, M@11568, M@11640, M@11684, M@11685, M@11729, M@11769, M@11813, M@11947, M@12378-1, M@12462, M@12765, M@13725, M@24470, M@4750028851, M@4750028853, M@5348B, M@5807B, M@5874B, M@5894B, M@5932B, M@5941B, M@5963B, M@6041B, M@6062B, M@6229B, M@6354B, M@6477B, M@6485B, M@6546B, M@6555B, M@6692B, M@6754B, M@6809B, M@6815B, M@6948B, M@7007B, M@7023B, M@7093B, M@7194B, M@7198B, M@7207B, M@7214B, M@7280B, M@7285B, M@7286B, M@7408B, M@7540B, M@7596B, M@7675B, M@7704B, M@8156B, M@8324B, M@8344B, M@8429B, M@8523B, M@8578B, M@8588B, M@8637B, M@8639RB, M@8648B, M@8808B, M@8813B, M@8898B, M@8977B, M@9050B, M@9261B, M@9276B, M@9355RB, M@9475B, M@9544B, M@9647B, M@9656B, M@9733B & M@9796B. ADAC ARGUS: Model #88295; Serial #: 9211014, 9301005, 9304019, 9304024, 9304028, 9401028, 9401031, 9401037, 9505014, 9910033, 93-10678- 1292, 9910028P, M@10225AMT, M@3732B, M@3903B, M@3954B, M@3956B, M@3960B, M@4249B, M@4415B, M@4646B, M@4868B, M@5172B, M@5321B, M@7341B & M@AMT1154. ADAC CIRRUS: Model #88294; 34, 718558, 9206009, 9207003, 9207004, 9207005, 9207007, 9211006, 9300113, 9304013, 9309012, 9404022, 9407028, 9509042, 9604098, 9910030, 93-10735-0393, 9621RQ, M@10110AMT, M@10324AMT, M@10402AMT, M@3599G-08, M@3905Q, M@4029Q, M@4129Q, M@4142Q, M@4668Q, M@4743Q, M@5003Q, M@5668Q, M@5895Q, M@AMT126R, M@AMT1282 & M@MVS3227A.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH recalled?

Yes. FDA published recall Z-0785-2014 on January 29, 2014. Reported hazard: The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.

What should I do if I have this product?