Back to search
FDAFDA · Food and drugs·Recall number Z-0787-2024

iQ Waste Well Adapter, Part Number: 700-3393

Published January 31, 2024 · Recall initiated: December 19, 2023

Summary

FDA records medical device Class II recall Z-0787-2024, published January 31, 2024 and initiated December 19, 2023. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 2 units; Lot: UDI/DI 15099590706180, all lot numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter, Inc.
Distributor
US and Australia, Brazil, Canada, China, France, Germany, Ghana, India, Israel, Italy, Lebanon, Lithuania, Mexico, Panama, Qatar, Romania, Saudi Arabia, Spain, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland,
Lot
UDI/DI 15099590706180, all lot numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was iQ Waste Well Adapter, Part Number: 700-3393 recalled?

Yes. FDA published recall Z-0787-2024 on January 31, 2024. Reported hazard: Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter