FDAFDA · Food and drugs·Recall number Z-0790-2013
Oxoid Antimicrobial Susceptibility Test Discs, Aztreonam 30 ug, IVD, REF CT0264B, Oxoid Ltd., UK, oxoid@thrmofisher.com (Both Remel & Oxoid, Ltd are a part of Thermo Fisher Scientific) An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure, by a disc-agar diffusion technique or a disc-broth elution technique, the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents (Aztreonam).
Published February 13, 2013 · Recall initiated: August 21, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm is recalling the product due to some cartridges may contain individual discs that are not impregnated with antibiotic. The firm notified their customers of the recall with a notification letter and requested a confirmation fax back of receipt.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Remel Inc
- Distributor
- Nationwide Distribution including WI, MO, MA, CT, AL, IL, WA, CA, RI, MD, NJ, NC and TX.
- Lot
- Lot 1100699
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.