ARCHITECT Free T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of free triiodothyronine (Free T3) in human serum and plasma.
Published February 13, 2019 · Recall initiated: October 12, 2018
Summary
FDA records medical device Class II recall Z-0798-2019, published February 13, 2019 and initiated October 12, 2018. Recalling firm: Abbott Ireland Diagnostics Division; Status: Terminated; Reported quantity: 191,410 units total; Lot: List number (Kit Configuration): 7K63-20 (4 x 100 tests/kit), 7K63 25 (1 x 100 tests/kit), 7K63 27 (1 x 100 tests/kit), 7K63 30 (4 x 500 tests/kit), 7K63 35 (1 x 500 tests/kit), 7K63 32 (4 x 500 tests/kit), 7K63 37 (1 x 500 tests/kit). All lot codes for each of these list numbers are included in the recall. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Samples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITECT i1000SR and i2000/i2999SR platforms.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Ireland Diagnostics Division
- Distributor
- Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR. International distribution to AFGHANISTAN, ALBANIA, ALGERIA, ANGOLA, ARMENIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA AND HERZEGOVI, BRUNEI, BULGARIA, BURKINA FASO, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GAZA & JERICHO, GEORGIA, GERMANY, GHANA, GREECE, HONDURAS, HUNGARY, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAWI, MALAYSIA, MALDIVES, MALI, MAURITANIA, MAURITIUS, MOLDOVA, MONTENEGRO, MOROCCO, MOZAMBIQUE, NEPAL, NETHERLANDS, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA,SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, ST. LUCIA, ST. VINCENT, SWEDEN, SWITZERLAND, TANZANIA, THE GAMBIA,TRINIDAD & TOBAGO, TUNISIA, TURKEY, UAE, UGANDA, UKRAINE,UNITED KINGDOM, VIETNAM, ZIMBABWE.
- Lot
- List number (Kit Configuration): 7K63-20 (4 x 100 tests/kit), 7K63 25 (1 x 100 tests/kit), 7K63 27 (1 x 100 tests/kit), 7K63 30 (4 x 500 tests/kit), 7K63 35 (1 x 500 tests/kit), 7K63 32 (4 x 500 tests/kit), 7K63 37 (1 x 500 tests/kit). All lot codes for each of these list numbers are included in the recall.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDAAscension Modular Radial Head (MRH) Radial Head Implant, Catalog Number, MRH-350-04-WWFebruary 13, 2019
- FDARichard Wolf Bronchoscope Tubes, Model Texas. Device is reusable and should be sterilized prior to each use.February 13, 2019
- FDAGE Healthcare Centricity Universal Viewer Breast Imaging Centricity Universal Viewer is a device that displaysFebruary 13, 2019
- FDAAguja de Aspiracion con Profundidad Ajustable Tipo I 16ga x 5 cm- Product Usage: Manual surgical instrumentsFebruary 13, 2019
- FDAAlinity i Progesterone Reagent Kit, List Number 08P3620January 2, 2019
- FDAARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 or 4 Bottle(s) (6.6 mL/27.0 mL) anti-T3 (sheep) coatedDecember 3, 2014
Questions about this recall
Was ARCHITECT Free T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of free triiodothyronine (Free T3) in human serum and plasma. recalled?
Yes. FDA published recall Z-0798-2019 on February 13, 2019. Reported hazard: Samples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITECT i1000SR and i2000/i2999SR platforms.
What should I do if I have this product?
Voluntary: Firm initiated · Letter