Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972
Published January 15, 2025 · Recall initiated: November 26, 2024
Summary
FDA records medical device Class II recall Z-0803-2025, published January 15, 2025 and initiated November 26, 2024. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 362 units; Lot: UDI: (01)15099590201937(17)241130(11)231201; Lot 339071. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the country of Canada.
- Lot
- UDI: (01)15099590201937(17)241130(11)231201; Lot 339071
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDAROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDAIQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345,October 21, 2026
- FDADxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513October 21, 2026
- FDAGreat Value, Chicken Broth, 48 oz. , Aseptic paper cartons, 6 retail units per case, 120 cases/pallet .January 15, 2025
- FDAThe FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurementJanuary 15, 2025
Questions about this recall
Was Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972 recalled?
Yes. FDA published recall Z-0803-2025 on January 15, 2025. Reported hazard: Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.
What should I do if I have this product?
Voluntary: Firm initiated · Letter