Video Monitor Suspensions that may be on the following Fluoroscopic and Radiographic X-Ray systems: Precision 500D, Legacy, Prestilix, Monitrol, RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII Product Usage: The systems are intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations on both adult and pediatric subjects. The systmes are used for taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
Published January 29, 2014 · Recall initiated: November 8, 2013
Summary
FDA records medical device Class II recall Z-0804-2014, published January 29, 2014 and initiated November 8, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 5651; Lot: serial numbers: 182, 00000000159728, 719723, 00000000045GP5, 00000000047GP1, 00000000120GP6, 00000000167GP7, 00000000168GP5, 00000000169GP3, 00000000170GP1, 00000000171GP9, 00000000175GP0, 00000000177GP6, 00000000180GP0, 00000000182GP6, 00000000184GP2, 00000000185GP9, 00000000186GP7, 00000000190GP9, 00000000223GQ8, 00000000252GP7, 00000000254GP3, 00000000256GP8, 00000000259GP2, 00000000277GQ4, 00000000331GP9, 00000000334GP3, 00000000353GP3, 00000000379GP8, 00000000380GP6, 00000000381GP4, 00000000382GP2, 00000000438GP2, 00000000439GP0, 00000000440GP8, 00000000444GP0, 00000000490GP3, 00000000491GP1, 00000000506GP6, 00000000556GP1, 00000000558GP7, 00000000590GP0, 00000000591GP8, 00000000592GP6, 00000000593GP4, 00000000594GP2, 00000000669GQ2,, 00000000694GP0, 00000000695GP7, 00000000701GP3, 00000000708GP8, 00000000778GP1, 00000000817GP7, 00000000818GP5, 00000000848GP2, 00000000884GP7, 00000000885GP4, 00000000917GP5, 00000000920GP9, 00000000921GP7, 00000000922GP5, 00000000923GP3, 00000000924GP1, 00000000946GP4, 00000000963GP9, 00000000964GP7, 00000000965GP4, 00000000AD6037, 00000000AD7002, 00000000AI6036, 00000001032GP2, 00000001033GP0, 00000001034GP8, 00000001035GP5, 00000001046GP2, 00000001067GP8, 00000001088GP4, 00000001091GP8, 00000001098GP3, 00000001106GP4, 00000001107GP2, 00000001109GP8, 00000001111GP4, 00000001232GP8, 00000001233GP6, 00000001267GQ4, 00000001341GP7, 00000001367GP2, 00000001374GP8, 00000001380M35, 00000001425GP8, 00000001433GP2, 00000001434GP0, 00000001435GP7, 00000001438GP1, 00000001441GP5, 00000001442GP3, 00000001453GP0, 00000001483GP7, 00000001540GP4, 00000001541GP2, 00000001543GP8, 00000001612GP1, 00000001614GP7, 00000001615GP4, 00000001663GP4, 00000001665GP9, 00000001737GP6, 00000001803M36, 00000001835GP8, 00000001848GP1, 00000001867GP1, 00000001868GP9, 00000001881M32, 00000001894GP5, 00000001895GP2, 00000001896GP0, 00000001897GP8, 00000001898GP6, 00000001942GP2, 00000001943GP0, 00000001944GP8, 00000001946GP3, 00000001971M31, 00000001991GP9, 00000001998GP4, 00000001999GP2, 00000002000GP8, 00000002115GP4, 00000002117GP0, 00000002194GP9, 00000002197GP2, 00000002198GP0, 00000002316GP8, 00000002319M32, 00000002337GP4, 00000002338GP2, 00000002340GP8, 00000002366GP3, 00000002509M38, 00000002530GP4, 00000002557GP7, 00000002562GP7, 00000002569GP2, 00000002570GP0, 00000002609GQ6, 00000002615M33, 00000002624M35, 00000002635GP1, 00000002636GP9, 00000002637GP7, 00000002639GP3, 00000002669GP0, 00000002839GP9, 00000002840GP7, 00000002894GP4, 00000002922M33, 00000002997GP5, 00000003189GP8, 00000003190GP6, 00000003191GP4, 00000003240GP9, 00000003241GP7, 00000003244GP1, 00000003337GP3, 00000003338GP1, 00000003339GP9, 00000003341GP5, 00000003343GP1, 00000003431GP4, 00000003442GP1, 00000003444GP7, 00000003458GP7, 00000003518GP8, 00000003519GP6, 00000003520GP4, 00000003583GP2, 00000003619GP4, 00000003621GP0, 00000003622GP8, 00000003671GQ5, 00000003718GP4, 00000003719GP2, 00000003720GP0, 00000003721GP8, 00000003724GP2, 00000003726GP7, 00000003727GP5, 00000003802GP6, 00000003804GP2, 00000003929GP7, 00000003967GP7, 00000003968GP5, 00000003970GP1, 00000003972GP7, 00000003973GP5, 00000003976GP8, 00000004029GP5, 00000004030GP3, 00000004033GP7, 00000004034GP5, 00000004035GP2, 00000004036GP0, 00000004037GP8, 00000004038GP6, 00000004108GQ7, 00000004182M32, 00000004223GP4, 00000004224GP2, 00000004226GP7, 00000004271GP3, 00000004272GP1, 00000004273GP9, 00000004274GP7, 00000004275GP4, 00000004278GQ8, 00000004279GQ6, 00000004300GP0, 00000004301GP8, 00000004302GP6, 00000004303GP4, 00000004336GP4, 00000004396GP8, 00000004398GP4, 00000004478GP4, 00000004480GP0, 00000004481GP8, 00000004485GP9, 00000004486GP7, 00000004539M33, 00000004540M31, 00000004631GP8, 00000004662GP3, 00000004663GP1, 00000004664GP9, 00000004667GP2, 00000004668GP0, 00000004670GP6, 00000004671GP4, 00000004758GP9, 00000004761GP3, 00000004762GP1, 00000004763GP9, 00000004764GP7, 00000004765GP4, 00000004792GP8, 00000004794GP4, 00000004795GP1, 00000004796GP9, 00000004801GP7, 0. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Missing screws in the overhead Video Monitor Suspension. GE discovered that the 4 outer screws that hold the overhead Video Monitor Suspension to the carriage were missing since installation of the system. GE is not aware of any fall of the overhead monitor suspension.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution - US Nationwide including Washington D.C., Guam, and Puerto Rico. Outside the US to include: Afghanistan, Algeria, Argentina, Australia, Austria, Azerbaijan, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Lebanon, Libyan Arab Jamahiriya, Macedonia, Malaysia, Mali, Malta, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Niger, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Yemen.
- Lot
- serial numbers: 182, 00000000159728, 719723, 00000000045GP5, 00000000047GP1, 00000000120GP6, 00000000167GP7, 00000000168GP5, 00000000169GP3, 00000000170GP1, 00000000171GP9, 00000000175GP0, 00000000177GP6, 00000000180GP0, 00000000182GP6, 00000000184GP2, 00000000185GP9, 00000000186GP7, 00000000190GP9, 00000000223GQ8, 00000000252GP7, 00000000254GP3, 00000000256GP8, 00000000259GP2, 00000000277GQ4, 00000000331GP9, 00000000334GP3, 00000000353GP3, 00000000379GP8, 00000000380GP6, 00000000381GP4, 00000000382GP2, 00000000438GP2, 00000000439GP0, 00000000440GP8, 00000000444GP0, 00000000490GP3, 00000000491GP1, 00000000506GP6, 00000000556GP1, 00000000558GP7, 00000000590GP0, 00000000591GP8, 00000000592GP6, 00000000593GP4, 00000000594GP2, 00000000669GQ2,, 00000000694GP0, 00000000695GP7, 00000000701GP3, 00000000708GP8, 00000000778GP1, 00000000817GP7, 00000000818GP5, 00000000848GP2, 00000000884GP7, 00000000885GP4, 00000000917GP5, 00000000920GP9, 00000000921GP7, 00000000922GP5, 00000000923GP3, 00000000924GP1, 00000000946GP4, 00000000963GP9, 00000000964GP7, 00000000965GP4, 00000000AD6037, 00000000AD7002, 00000000AI6036, 00000001032GP2, 00000001033GP0, 00000001034GP8, 00000001035GP5, 00000001046GP2, 00000001067GP8, 00000001088GP4, 00000001091GP8, 00000001098GP3, 00000001106GP4, 00000001107GP2, 00000001109GP8, 00000001111GP4, 00000001232GP8, 00000001233GP6, 00000001267GQ4, 00000001341GP7, 00000001367GP2, 00000001374GP8, 00000001380M35, 00000001425GP8, 00000001433GP2, 00000001434GP0, 00000001435GP7, 00000001438GP1, 00000001441GP5, 00000001442GP3, 00000001453GP0, 00000001483GP7, 00000001540GP4, 00000001541GP2, 00000001543GP8, 00000001612GP1, 00000001614GP7, 00000001615GP4, 00000001663GP4, 00000001665GP9, 00000001737GP6, 00000001803M36, 00000001835GP8, 00000001848GP1, 00000001867GP1, 00000001868GP9, 00000001881M32, 00000001894GP5, 00000001895GP2, 00000001896GP0, 00000001897GP8, 00000001898GP6, 00000001942GP2, 00000001943GP0, 00000001944GP8, 00000001946GP3, 00000001971M31, 00000001991GP9, 00000001998GP4, 00000001999GP2, 00000002000GP8, 00000002115GP4, 00000002117GP0, 00000002194GP9, 00000002197GP2, 00000002198GP0, 00000002316GP8, 00000002319M32, 00000002337GP4, 00000002338GP2, 00000002340GP8, 00000002366GP3, 00000002509M38, 00000002530GP4, 00000002557GP7, 00000002562GP7, 00000002569GP2, 00000002570GP0, 00000002609GQ6, 00000002615M33, 00000002624M35, 00000002635GP1, 00000002636GP9, 00000002637GP7, 00000002639GP3, 00000002669GP0, 00000002839GP9, 00000002840GP7, 00000002894GP4, 00000002922M33, 00000002997GP5, 00000003189GP8, 00000003190GP6, 00000003191GP4, 00000003240GP9, 00000003241GP7, 00000003244GP1, 00000003337GP3, 00000003338GP1, 00000003339GP9, 00000003341GP5, 00000003343GP1, 00000003431GP4, 00000003442GP1, 00000003444GP7, 00000003458GP7, 00000003518GP8, 00000003519GP6, 00000003520GP4, 00000003583GP2, 00000003619GP4, 00000003621GP0, 00000003622GP8, 00000003671GQ5, 00000003718GP4, 00000003719GP2, 00000003720GP0, 00000003721GP8, 00000003724GP2, 00000003726GP7, 00000003727GP5, 00000003802GP6, 00000003804GP2, 00000003929GP7, 00000003967GP7, 00000003968GP5, 00000003970GP1, 00000003972GP7, 00000003973GP5, 00000003976GP8, 00000004029GP5, 00000004030GP3, 00000004033GP7, 00000004034GP5, 00000004035GP2, 00000004036GP0, 00000004037GP8, 00000004038GP6, 00000004108GQ7, 00000004182M32, 00000004223GP4, 00000004224GP2, 00000004226GP7, 00000004271GP3, 00000004272GP1, 00000004273GP9, 00000004274GP7, 00000004275GP4, 00000004278GQ8, 00000004279GQ6, 00000004300GP0, 00000004301GP8, 00000004302GP6, 00000004303GP4, 00000004336GP4, 00000004396GP8, 00000004398GP4, 00000004478GP4, 00000004480GP0, 00000004481GP8, 00000004485GP9, 00000004486GP7, 00000004539M33, 00000004540M31, 00000004631GP8, 00000004662GP3, 00000004663GP1, 00000004664GP9, 00000004667GP2, 00000004668GP0, 00000004670GP6, 00000004671GP4, 00000004758GP9, 00000004761GP3, 00000004762GP1, 00000004763GP9, 00000004764GP7, 00000004765GP4, 00000004792GP8, 00000004794GP4, 00000004795GP1, 00000004796GP9, 00000004801GP7, 0
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDACivco Belly Board MRI, Reusable non-sterile glass fiber belly board with cushion, REF 126000, packaged 1 perJanuary 29, 2014
- FDAORCHESTRA/ORCHESTRA PLUS ProgrammerJanuary 29, 2014
Questions about this recall
Was Video Monitor Suspensions that may be on the following Fluoroscopic and Radiographic X-Ray systems: Precision 500D, Legacy, Prestilix, Monitrol, RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII Product Usage: The systems are intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations on both adult and pediatric subjects. The systmes are used for taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. recalled?
Yes. FDA published recall Z-0804-2014 on January 29, 2014. Reported hazard: Missing screws in the overhead Video Monitor Suspension. GE discovered that the 4 outer screws that hold the overhead Video Monitor Suspension to the carriage were missing since installation of the system. GE is not aware of any fall of the overhead monitor suspension.
What should I do if I have this product?
Voluntary: Firm initiated · Letter