Cloth electrodes with Carbon construction: P/N 201-1231, Cat Nos.: 201-123, 202-454, 203-118, 203-342, 203-454, 203-493, 203-672, 400-877, and 400-877-BJC, 2" x 2" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1241, Cat Nos.: 201-124, 202-455, 203-340, 203-674, and 400-872, 2" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1251, Cat Nos.: 201-138, 202-276, 202-456, 203-336, 203-680, and 400-879, 3" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1381, Cat Nos.: 201-126, 203-678, and 400-856, 2" x 3.5" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1501, Cat Nos.: 201-150 and 400-863, 1.25" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1521, Cat No. 400-852, 3 x 5 Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1541, Cat Nos.:201-154 and 400-865 1.5 x 2.5 Oval Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 Cutaneous electrode.
Published February 20, 2013 · Recall initiated: November 30, 2009
Summary
FDA records medical device Class II recall Z-0815-2013, published February 20, 2013 and initiated November 30, 2009. Recalling firm: NAImco Inc dba Richmar Inc; Status: Terminated; Reported quantity: 209860 electrodes; Lot: All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- NAImco Inc dba Richmar Inc
- Distributor
- Distributed in the states of GA and TN.
- Lot
- All Serial Numbers distributed between 11/01/2009 and 01/31/2010
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFoam Electrodes with Carbon construction: P/N 201-1261, Cat Nos.: 201-126, 203-678, and 400-856, 1.5"February 20, 2013
- FDAFoam electrodes with Silver-Carbon construction: P/N 202-9271, Cat No. 400-898, 2" x 2"February 20, 2013
- FDACloth electrodes with Silver-Carbon construction: P/N 202-9231, Cat No. 400-899, 2" x 2"February 20, 2013
- FDAGE Healthcare, Dash Port 2 Docking Station. Product Usage: is a docking station (quick mount/dismount base)February 20, 2013
- FDAGE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Physiological Patient Monitor and accessories.February 20, 2013
- FDAUltradent Products, Chromaclone PVS, Fast Set heavy body, Bubble Gum Flavor, 50 ml, Impression Material, PartFebruary 20, 2013
Questions about this recall
Was Cloth electrodes with Carbon construction: P/N 201-1231, Cat Nos.: 201-123, 202-454, 203-118, 203-342, 203-454, 203-493, 203-672, 400-877, and 400-877-BJC, 2" x 2" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1241, Cat Nos.: 201-124, 202-455, 203-340, 203-674, and 400-872, 2" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1251, Cat Nos.: 201-138, 202-276, 202-456, 203-336, 203-680, and 400-879, 3" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1381, Cat Nos.: 201-126, 203-678, and 400-856, 2" x 3.5" Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1501, Cat Nos.: 201-150 and 400-863, 1.25" Round Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1521, Cat No. 400-852, 3 x 5 Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1541, Cat Nos.:201-154 and 400-865 1.5 x 2.5 Oval Electrode, Cloth, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 Cutaneous electrode. recalled?
Yes. FDA published recall Z-0815-2013 on February 20, 2013. Reported hazard: Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
What should I do if I have this product?
Voluntary: Firm initiated · Letter