Foam Electrodes with Carbon construction: P/N 201-1261, Cat Nos.: 201-126, 203-678, and 400-856, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1271, Cat Nos.: 201-127, 202-093, 203-483, 203-565, 203-673, and 400-873, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1281, Cat Nos.: 201-128, 202-277, 203-675, and 400-873, 2" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1291, Cat Nos.: 201-129, 203-677, and 400-882, 3" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1301, Cat Nos.: 201-130, 203-679, and 400-855, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1371, Cat Nos.: 201-137, 202-278, 203-484, 203-681, and 400-880, 2" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-5671, Cat No. 202-567, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Published February 20, 2013 · Recall initiated: November 30, 2009

Summary

FDA records medical device Class II recall Z-0817-2013, published February 20, 2013 and initiated November 30, 2009. Recalling firm: NAImco Inc dba Richmar Inc; Status: Terminated; Reported quantity: 129400 electrodes; Lot: All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
NAImco Inc dba Richmar Inc
Distributor
Distributed in the states of GA and TN.
Lot
All Serial Numbers distributed between 11/01/2009 and 01/31/2010
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Foam Electrodes with Carbon construction: P/N 201-1261, Cat Nos.: 201-126, 203-678, and 400-856, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1271, Cat Nos.: 201-127, 202-093, 203-483, 203-565, 203-673, and 400-873, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1281, Cat Nos.: 201-128, 202-277, 203-675, and 400-873, 2" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1291, Cat Nos.: 201-129, 203-677, and 400-882, 3" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1301, Cat Nos.: 201-130, 203-679, and 400-855, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1371, Cat Nos.: 201-137, 202-278, 203-484, 203-681, and 400-880, 2" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-5671, Cat No. 202-567, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode. recalled?