ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br®nsh®j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.
Published January 29, 2014 · Recall initiated: October 15, 2013
Summary
FDA records medical device Class II recall Z-0816-2014, published January 29, 2014 and initiated October 15, 2013. Recalling firm: Radiometer America Inc; Status: Terminated; Reported quantity: Total 3414 (US 524, CA 31, ROW 2859); Lot: Lot R0375N009, Model #393-090, S/N 393-090R0027N001 - 393-090R0379N010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Radiometer America Inc
- Distributor
- Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI & WV and Internationally to Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, Italy, Japan, Netherlands, Norway, Poland, Russia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey & United Kingdom.
- Lot
- Lot R0375N009, Model #393-090, S/N 393-090R0027N001 - 393-090R0379N010
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAABL800 analyzer, model numbers 393-800 and 393-801.January 2, 2019
- FDAAutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+October 11, 2017
- FDAABL90 Flex Model #: 393-090 Portable, automated analyzer that measures pH, blood gases, electrolytes, glucose,March 22, 2017
- FDATCM CombiM Modules 903-111May 11, 2016
- FDATitanium Single Lumen Low-Profile port, with Pre-Attached open-Ended Silicone 6.6Fr Catheter. Product CodeJanuary 29, 2014
- FDAVITROS Chemistry Products COCM ReagentJanuary 29, 2014
Questions about this recall
Was ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br®nsh®j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. recalled?
Yes. FDA published recall Z-0816-2014 on January 29, 2014. Reported hazard: RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter