ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br®nsh®j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.

Published January 29, 2014 · Recall initiated: October 15, 2013

Summary

FDA records medical device Class II recall Z-0816-2014, published January 29, 2014 and initiated October 15, 2013. Recalling firm: Radiometer America Inc; Status: Terminated; Reported quantity: Total 3414 (US 524, CA 31, ROW 2859); Lot: Lot R0375N009, Model #393-090, S/N 393-090R0027N001 - 393-090R0379N010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Radiometer America Inc
Distributor
Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI & WV and Internationally to Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, Italy, Japan, Netherlands, Norway, Poland, Russia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey & United Kingdom.
Lot
Lot R0375N009, Model #393-090, S/N 393-090R0027N001 - 393-090R0379N010
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br®nsh®j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. recalled?

Yes. FDA published recall Z-0816-2014 on January 29, 2014. Reported hazard: RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter