Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
Published January 29, 2020 · Recall initiated: December 16, 2019
Summary
FDA records medical device Class II recall Z-0819-2020, published January 29, 2020 and initiated December 16, 2019. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 205; Lot: Model 2088026-9XX Versions 6.0 SP10 or higher. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
- Lot
- Model 2088026-9XX Versions 6.0 SP10 or higher
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAPiccadilly Cornbread Dressing in 2# traysJanuary 29, 2020
- FDAKoios DS Breast <series 1.0>. The device is a software application which assists skilled physicians inJanuary 29, 2020
Questions about this recall
Was Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources. recalled?
Yes. FDA published recall Z-0819-2020 on January 29, 2020. Reported hazard: Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
What should I do if I have this product?
Voluntary: Firm initiated · Letter