Back to search
FDAFDA · Food and drugs·Recall number Z-0819-2020

Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.

Published January 29, 2020 · Recall initiated: December 16, 2019

Summary

FDA records medical device Class II recall Z-0819-2020, published January 29, 2020 and initiated December 16, 2019. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 205; Lot: Model 2088026-9XX Versions 6.0 SP10 or higher. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
Lot
Model 2088026-9XX Versions 6.0 SP10 or higher
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources. recalled?

Yes. FDA published recall Z-0819-2020 on January 29, 2020. Reported hazard: Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.

What should I do if I have this product?

Voluntary: Firm initiated · Letter