Viterion TeleHealth Monitor Model Number V500; Product Codes 08832984 (US) and 02080468 (UK) - Physiological patient monitor (without arrhythmia detection or alarms). Viterion V100, V100-BGM and V500 Telehealth Monitors transfer medical information.
Published February 20, 2013 · Recall initiated: November 15, 2012
Summary
FDA records medical device Class II recall Z-0822-2013, published February 20, 2013 and initiated November 15, 2012. Recalling firm: Viterion TeleHealthcare Llc; Status: Terminated; Reported quantity: Domestic - 296 units, Foreign - 73 units; Lot: All units are affected. Serial Numbers A3303HA000201, A3303HA000202, A3303HA000203, A3303HA000204, A3303HA000205, A3304HA000206, A3304HA000207, A3304HA000208, A3304HA000209, A3304HA000210, A3305HA000211, A3305HA000212, A3305HA000213, A3305HA000214, A3305HA000215, A3306HA000216, A3306HA000217, A3306HA000218, A3306HA000219, A3306HA000220, A3310HA000221, A3310HA000222, A3310HA000223, A3310HA000224, A3310HA000225, A3404HA000226, A3404HA000227, A3404HA000228, A3404HA000229, A3404HA000230, A3404HA000231, A3404HA000232, A3404HA000233, A3404HA000234, A3404HA000235, A3404HA000236, A3404HA000237, A3404HA000238, A3404HA000239, A3404HA000240, A3404HA000241, A3404HA000242, A3404HA000243, A3404HA000244, A3404HA000245, A3404HA000246, A3404HA000247, A3404HA000248, A3404HA000249, A3404HA000250, A3602HA000252, A3602HA000253, A3602HA000254, A3602HA000255, A3602HA000256, A3602HA000257, A3602HA000258, A3602HA000259, A3602HA000260, A3602HA000261, A3602HA000262, A3602HA000263, A3602HA000264, A3602HA000265, A3602HA000266, A3701HA000267, A3701HA000268, A3701HA000269, A3701HA000270, A3701HA000271, A3701HA000272, A3701HA000273, A3701HA000274, A3701HA000275, A3701HA000276, A3701HA000277, A3701HA000278, A3701HA000279, A3701HA000280, A3701HA000281, A3701HA000282, A3701HA000283, A3701HA000284, A3701HA000285, A3701HA000286, A3701HA000287, A3701HA000288, A3701HA000289, A3701HA000290, A3701HA000291, A3818HA000292, A3818HA000293, A3818HA000294, A3818HA000295, A3818HA000296, A3818HA000297, A3818HA000298, A3818HA000299, A3818HA000300, A3818HA000301, A3818HA000302, A3818HA000303, A3818HA000304, A3818HA000305, A3818HA000306, A3818HA000307, A3818HA000308, A3818HA000309, A3818HA000310, A3818HA000311, A3818HA000312, A3818HA000313, A3818HA000314, A3818HA000315, A3818HA000316, A3905HA000317, A3905HA000318, A3905HA000319, A3905HA000320, A3905HA000321, A3905HA000322, A3905HA000323, A3905HA000324, A3905HA000325, A3905HA000326, A3905HA000327, A3905HA000328, A3905HA000329, A3905HA000330, A3905HA000331, A3905HA000332, A3905HA000333, A3905HA000334, A3905HA000335, A3905HA000336, A3905HA000337, A3905HA000338, A3905HA000339, A3905HA000340, A3905HA000341, A3D01HA000342, A3D01HA000343, A3D01HA000344, A3D01HA000345, A3D01HA000346, A3D01HA000347, A3D01HA000348, A3D01HA000349, A3D01HA000350, A3D01HA000351, A3D01HA000352, A3D01HA000353, A3D01HA000354, A3D01HA000355, A3D01HA000356, A3D01HA000357, A3D01HA000358, A3D01HA000359, A3D01HA000360, A3D01HA000361, A3D01HA000362, A3D01HA000363, A3D01HA000364, A3D01HA000365, A3D01HA000366, A3D01HA000367, A3D01HA000368, A3D01HA000369, A3D01HA000370, A3D01HA000371, A3D01HA000372, A3D01HA000373, A3D01HA000374, A3D01HA000375, A3D01HA000376, A3D01HA000377, A3D01HA000378, A3D01HA000379, A3D01HA000380, A3D01HA000381, A3D01HA000382, A3D01HA000383, A3D01HA000384, A3D01HA000385, A3D01HA000386, A3D01HA000387, A3D01HA000388, A3D01HA000389, A3D01HA000390, A4113HA000391, A4113HA000392, A4113HA000393, A4113HA000394, A4113HA000395, A4113HA000396, A4113HA000397, A4113HA000398, A4113HA000399, A4113HA000400, A4113HA000401, A4113HA000402, A4113HA000403, A4113HA000404, A4113HA000405, A4113HA000406, A4113HA000407, A4113HA000408, A4113HA000409, A4113HA000410, A4113HA000411, A4113HA000412, A4113HA000413, A4113HA000414, A4113HA000415, A4113HA000416, A4113HA000417, A4113HA000418, A4113HA000419, A4113HA000420, A4113HA000421, A4113HA000422, A4113HA000423, A4113HA000424, A4113HA000425, A4113HA000426, A4113HA000427, A4113HA000428, A4113HA000429, A4113HA000430, A4113HA000431, A4113HA000432, A4113HA000433, A4113HA000434, A4113HA000435, A4113HA000436, A4113HA000437, A4113HA000438, A4113HA000439, A4113HA000440, A4113HA000441, A4526MA000442, A4526MA000443, A4526MA000444, A4526MA000446, A4712MA000445, A4712MA000447, A4712MA000448, A4712MA000449, A4712MA000450, A4712MA000451, A4712MA000452, A4712MA000453, A4712MA000454, A4712MA000455, A4712MA000456, A4712MA000457, A4712MA000458, A4712MA000459, A4712MA000460, A4712MA000461, A4712MA000462, A4712MA000463, A4712MA000464, A4712MA000465, . Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04 meters to healthcare professionals.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Viterion TeleHealthcare Llc
- Distributor
- Worldwide Distribution-USA (nationwide) and the United Kingdom.
- Lot
- All units are affected. Serial Numbers A3303HA000201, A3303HA000202, A3303HA000203, A3303HA000204, A3303HA000205, A3304HA000206, A3304HA000207, A3304HA000208, A3304HA000209, A3304HA000210, A3305HA000211, A3305HA000212, A3305HA000213, A3305HA000214, A3305HA000215, A3306HA000216, A3306HA000217, A3306HA000218, A3306HA000219, A3306HA000220, A3310HA000221, A3310HA000222, A3310HA000223, A3310HA000224, A3310HA000225, A3404HA000226, A3404HA000227, A3404HA000228, A3404HA000229, A3404HA000230, A3404HA000231, A3404HA000232, A3404HA000233, A3404HA000234, A3404HA000235, A3404HA000236, A3404HA000237, A3404HA000238, A3404HA000239, A3404HA000240, A3404HA000241, A3404HA000242, A3404HA000243, A3404HA000244, A3404HA000245, A3404HA000246, A3404HA000247, A3404HA000248, A3404HA000249, A3404HA000250, A3602HA000252, A3602HA000253, A3602HA000254, A3602HA000255, A3602HA000256, A3602HA000257, A3602HA000258, A3602HA000259, A3602HA000260, A3602HA000261, A3602HA000262, A3602HA000263, A3602HA000264, A3602HA000265, A3602HA000266, A3701HA000267, A3701HA000268, A3701HA000269, A3701HA000270, A3701HA000271, A3701HA000272, A3701HA000273, A3701HA000274, A3701HA000275, A3701HA000276, A3701HA000277, A3701HA000278, A3701HA000279, A3701HA000280, A3701HA000281, A3701HA000282, A3701HA000283, A3701HA000284, A3701HA000285, A3701HA000286, A3701HA000287, A3701HA000288, A3701HA000289, A3701HA000290, A3701HA000291, A3818HA000292, A3818HA000293, A3818HA000294, A3818HA000295, A3818HA000296, A3818HA000297, A3818HA000298, A3818HA000299, A3818HA000300, A3818HA000301, A3818HA000302, A3818HA000303, A3818HA000304, A3818HA000305, A3818HA000306, A3818HA000307, A3818HA000308, A3818HA000309, A3818HA000310, A3818HA000311, A3818HA000312, A3818HA000313, A3818HA000314, A3818HA000315, A3818HA000316, A3905HA000317, A3905HA000318, A3905HA000319, A3905HA000320, A3905HA000321, A3905HA000322, A3905HA000323, A3905HA000324, A3905HA000325, A3905HA000326, A3905HA000327, A3905HA000328, A3905HA000329, A3905HA000330, A3905HA000331, A3905HA000332, A3905HA000333, A3905HA000334, A3905HA000335, A3905HA000336, A3905HA000337, A3905HA000338, A3905HA000339, A3905HA000340, A3905HA000341, A3D01HA000342, A3D01HA000343, A3D01HA000344, A3D01HA000345, A3D01HA000346, A3D01HA000347, A3D01HA000348, A3D01HA000349, A3D01HA000350, A3D01HA000351, A3D01HA000352, A3D01HA000353, A3D01HA000354, A3D01HA000355, A3D01HA000356, A3D01HA000357, A3D01HA000358, A3D01HA000359, A3D01HA000360, A3D01HA000361, A3D01HA000362, A3D01HA000363, A3D01HA000364, A3D01HA000365, A3D01HA000366, A3D01HA000367, A3D01HA000368, A3D01HA000369, A3D01HA000370, A3D01HA000371, A3D01HA000372, A3D01HA000373, A3D01HA000374, A3D01HA000375, A3D01HA000376, A3D01HA000377, A3D01HA000378, A3D01HA000379, A3D01HA000380, A3D01HA000381, A3D01HA000382, A3D01HA000383, A3D01HA000384, A3D01HA000385, A3D01HA000386, A3D01HA000387, A3D01HA000388, A3D01HA000389, A3D01HA000390, A4113HA000391, A4113HA000392, A4113HA000393, A4113HA000394, A4113HA000395, A4113HA000396, A4113HA000397, A4113HA000398, A4113HA000399, A4113HA000400, A4113HA000401, A4113HA000402, A4113HA000403, A4113HA000404, A4113HA000405, A4113HA000406, A4113HA000407, A4113HA000408, A4113HA000409, A4113HA000410, A4113HA000411, A4113HA000412, A4113HA000413, A4113HA000414, A4113HA000415, A4113HA000416, A4113HA000417, A4113HA000418, A4113HA000419, A4113HA000420, A4113HA000421, A4113HA000422, A4113HA000423, A4113HA000424, A4113HA000425, A4113HA000426, A4113HA000427, A4113HA000428, A4113HA000429, A4113HA000430, A4113HA000431, A4113HA000432, A4113HA000433, A4113HA000434, A4113HA000435, A4113HA000436, A4113HA000437, A4113HA000438, A4113HA000439, A4113HA000440, A4113HA000441, A4526MA000442, A4526MA000443, A4526MA000444, A4526MA000446, A4712MA000445, A4712MA000447, A4712MA000448, A4712MA000449, A4712MA000450, A4712MA000451, A4712MA000452, A4712MA000453, A4712MA000454, A4712MA000455, A4712MA000456, A4712MA000457, A4712MA000458, A4712MA000459, A4712MA000460, A4712MA000461, A4712MA000462, A4712MA000463, A4712MA000464, A4712MA000465,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASLIMDIA Revolution capsules, 30-count capsules per bottle, Yerba Naturals, 19685 Varda Ave., Buena Park, CA.February 27, 2013
- FDAViterion TeleHealth Monitor Model Number V100-BGM, Product Code 81842478 (US) -- Physiological patient monitorFebruary 20, 2013
- FDAViterion TeleHealth Monitor Model Number V100, Product Codes 09377512 (US) and 04480218 (UK). PhysiologicalFebruary 20, 2013
- FDAGE Healthcare, CARESCAPE Monitor B650. The CARESCAPE Monitor B650 is a multi-parameter patient monitorFebruary 20, 2013
- FDAPhilips Healthcare SureSigns VS2+ Model 863278 SureSigns VS2+ NBP and SpO2 Model 863279 SureSigns VS2+ NBP,February 20, 2013
- FDAQuidel, Sofia Influenza A + B FIA, Kit #20218 (25-Test Kit), in vitro diagnostic.February 20, 2013
Questions about this recall
Was Viterion TeleHealth Monitor Model Number V500; Product Codes 08832984 (US) and 02080468 (UK) - Physiological patient monitor (without arrhythmia detection or alarms). Viterion V100, V100-BGM and V500 Telehealth Monitors transfer medical information. recalled?
Yes. FDA published recall Z-0822-2013 on February 20, 2013. Reported hazard: Viterion telehealth monitors do not transmit "HI" (above the meter's measurable range) and "LO" (below the meter's measurable range) blood glucose readings from Bayer's Contour Elite XL blood glucose meters, or from Roche's Accu-Chek Advantage 03 and 04 meters to healthcare professionals.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone