Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) Breast Plastics Pack, kit number BBPL20D; b) MINOR PLASTIC, kit number WPMP16H
Published February 7, 2024 · Recall initiated: November 29, 2023
Summary
FDA records medical device Class II recall Z-0824-2024, published February 7, 2024 and initiated November 29, 2023. Recalling firm: American Contract Systems, Inc.; Status: Ongoing; Reported quantity: 112 units; Lot: a) BBPL20D, Lot Codes: 862221, exp. 5/18/2023; 688221, exp. 5/31/2023; 756221, exp. 9/1/2023; 782221, exp. 8/6/2023; b) WPMP16H,Lot Codes: 904231, exp. 4/6/2024; 966231, exp. 2/3/2024; 650221, exp. 12/16/2023; 708221, exp. 10/19/2023; 805221, exp. 7/14/2023; 888221, exp. 4/22/2023; 867221, exp. 5/13/2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Contract Systems, Inc.
- Distributor
- US Nationwide distribution.
- Lot
- a) BBPL20D, Lot Codes: 862221, exp. 5/18/2023; 688221, exp. 5/31/2023; 756221, exp. 9/1/2023; 782221, exp. 8/6/2023; b) WPMP16H,Lot Codes: 904231, exp. 4/6/2024; 966231, exp. 2/3/2024; 650221, exp. 12/16/2023; 708221, exp. 10/19/2023; 805221, exp. 7/14/2023; 888221, exp. 4/22/2023; 867221, exp. 5/13/2023
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Banner University Med Ctr. Product Name: Cardiac Cath Lab Pack-170278 Model/Catalog Number:April 2, 2025
- FDAHand Pack, AKHD97C; Medical convenience kitOctober 2, 2024
- FDAVaginal Delivery, ASDV25B; Medical convenience kitOctober 2, 2024
- FDAIR Procedure Pack, CEAT19W; Medical convenience kitOctober 2, 2024
- FDACustom procedural convenience kits and trays, urological, labeled as: a) UROLOGY DAVINCI PACK, kit numberFebruary 7, 2024
- FDACustom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPYFebruary 7, 2024
Questions about this recall
Was Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) Breast Plastics Pack, kit number BBPL20D; b) MINOR PLASTIC, kit number WPMP16H recalled?
Yes. FDA published recall Z-0824-2024 on February 7, 2024. Reported hazard: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail