Back to search
FDAFDA · Food and drugs·Recall number Z-0825-2024

Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48

Published February 7, 2024 · Recall initiated: November 29, 2023

Summary

FDA records medical device Class II recall Z-0825-2024, published February 7, 2024 and initiated November 29, 2023. Recalling firm: American Contract Systems, Inc.; Status: Ongoing; Reported quantity: 7881 units; Lot: a) AGBR90E, UDI/DI 00191072036865, Lot Numbers: 959231, exp. 02/10/2024; 681221, exp. 11/15/2023; 721221, exp. 09/30/2023; 771221, exp. 08/147/2023; 827221, exp. 06/22/2023; 897221, exp. 04/13/2023; b) AGMM60M, UDI/DI 00191072063786, Lot Numbers: 939231, exp. 03/02/2024; 638221, exp. 4528812/28/2023; 694221, exp. 11/02/2023; 734221, exp. 09/23/2023; 769221, exp. 08/19/2023; c) AHPO55J, UDI/DI 00191072171566, Lot Numbers: 682221, exp. 11/14/2023; 661221, exp. 12/5/2023; 977231, exp. 1/23/2024; 945231, exp. 2/24/2024; 891231, exp. 4/19/2024; d) BBGP31C, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 738221, exp. 9/19/2023; 805221, exp. 7/14/2023; 870221, exp. 5/10/2023; e) FDFT75N, UDI/DI none, Lot Numbers: 773221, exp. 8/15/2023; 714221, exp. 10/13/2023; 990231, exp. 1/10/2024; 935231, exp. 3/6/2024; 890231, exp. 4/20/2024; 870231, exp. 5/10/2024; f) HCAR04K, UDI/DI 00191072114174, Lot Numbers: 757221, exp. 08/31/2023; g) HGBA92F, UDI/DI none, Lot Numbers: 757221, exp. 8/31/2023; 695221, exp. 11/1/2023; 635221, exp. 12/31/2023; h) HNLP12J, UDI/DI 00191072146274, Lot Numbers: 805221, exp. 7/14/2023; 742221, exp. 9/15/2023; 673221, exp. 11/23/2023; 924231, exp. 3/17/2024; i) HSMD20E, UDI/DI 00191072130068, Lot Numbers: 898221, exp. 4/12/2023; 772221, exp. 8/16/2023; 984231, exp. 1/16/2024; j) JSMI10I, UDI/DI none, Lot Numbers: 766221, exp. 08/22/2023; k) LVEV61, UDI/DI 00191072033475, Lot Numbers: 678221, exp. 11/18/2023; 637221, exp. 12/29/2023; l) LVMJ21, UDI/DI 00191072139023, Lot Numbers: 833221, exp. 6/16/2023; 764221, exp. 8/24/2023; 717221, exp. 10/10/2023; 988231, exp. 1/12/2024; 954231, exp. 2/15/2024; m) MMHN37F, UDI/DI none, Lot Numbers: 639221, exp. 12/27/2023; n) MMNU34E, UDI/DI none, Lot Numbers: 928231, exp. 3/13/2024; 889231, exp. 4/21/2024; o) MMUN28E, UDI/DI none, Lot Numbers: 889231, exp. 04/21/2024; p) OSVP16, UDI/DI 00191072156570, Lot Numbers: 827221, exp. 6/17/2023; q) OSVP16, UDI/DI 00191072170835, Lot Numbers: 700221, exp. 10/23/2023; 644221, exp. 12/4/2023; 987231, exp. 11/20/2023; r) SALM40H, UDI/DI none, Lot Numbers: 826221, exp. 6/23/2023; 888221, exp. 4/22/2023; s) SALM40I, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 756221, exp. 9/1/2023; 984231, exp. 1/16/2024; t) SAMB25E, UDI/DI none, Lot Numbers: 843221, exp. 6/6/2023; 894221, exp. 4/16/2023; 871221, exp. 5/9/2023; 686221, exp. 11/10/2023; 758221, exp. 8/30/2023; 793221, exp. 7/26/2023; 974231, exp. 1/26/2024; u) SAMJ36G, UDI/DI none, Lot Numbers: 660221, exp. 12/6/2023; 704221, exp. 10/23/2023; 752221, exp. 9/5/2023; 800221, exp. 7/19/2023; v) SMMN65, UDI/DI 00191072188472, Lot Numbers: 969231, exp. 1/31/2024; 926231, exp. 3/15/2024; w) SMMN65, UDI/DI none, Lot Numbers: 919231, exp. 3/22/2024; x) THGN59D, UDI/DI 00191072191069, Lot Numbers: 805221, exp. 7/14/2023; 721221, exp. 10/6/2023; y) TPMN15, UDI/DI 00191072171771, Lot Numbers: 737221, exp. 9/20/2023; 702221, exp. 10/25/2023; 664221, exp. 12/2/2023; 990231, exp. 1/10/2024; 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; 903231, exp. 4/7/2024; z) TPUN48, UDI/DI 00191072171832, Lot Numbers: 738221, exp. 9/19/2023; 688221, exp. 11/8/2023; 637221, exp. 12/29/2023; 967231, exp. 2/2/2024; 924231, exp. 3/17/2024. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
American Contract Systems, Inc.
Distributor
US Nationwide distribution.
Lot
a) AGBR90E, UDI/DI 00191072036865, Lot Numbers: 959231, exp. 02/10/2024; 681221, exp. 11/15/2023; 721221, exp. 09/30/2023; 771221, exp. 08/147/2023; 827221, exp. 06/22/2023; 897221, exp. 04/13/2023; b) AGMM60M, UDI/DI 00191072063786, Lot Numbers: 939231, exp. 03/02/2024; 638221, exp. 4528812/28/2023; 694221, exp. 11/02/2023; 734221, exp. 09/23/2023; 769221, exp. 08/19/2023; c) AHPO55J, UDI/DI 00191072171566, Lot Numbers: 682221, exp. 11/14/2023; 661221, exp. 12/5/2023; 977231, exp. 1/23/2024; 945231, exp. 2/24/2024; 891231, exp. 4/19/2024; d) BBGP31C, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 738221, exp. 9/19/2023; 805221, exp. 7/14/2023; 870221, exp. 5/10/2023; e) FDFT75N, UDI/DI none, Lot Numbers: 773221, exp. 8/15/2023; 714221, exp. 10/13/2023; 990231, exp. 1/10/2024; 935231, exp. 3/6/2024; 890231, exp. 4/20/2024; 870231, exp. 5/10/2024; f) HCAR04K, UDI/DI 00191072114174, Lot Numbers: 757221, exp. 08/31/2023; g) HGBA92F, UDI/DI none, Lot Numbers: 757221, exp. 8/31/2023; 695221, exp. 11/1/2023; 635221, exp. 12/31/2023; h) HNLP12J, UDI/DI 00191072146274, Lot Numbers: 805221, exp. 7/14/2023; 742221, exp. 9/15/2023; 673221, exp. 11/23/2023; 924231, exp. 3/17/2024; i) HSMD20E, UDI/DI 00191072130068, Lot Numbers: 898221, exp. 4/12/2023; 772221, exp. 8/16/2023; 984231, exp. 1/16/2024; j) JSMI10I, UDI/DI none, Lot Numbers: 766221, exp. 08/22/2023; k) LVEV61, UDI/DI 00191072033475, Lot Numbers: 678221, exp. 11/18/2023; 637221, exp. 12/29/2023; l) LVMJ21, UDI/DI 00191072139023, Lot Numbers: 833221, exp. 6/16/2023; 764221, exp. 8/24/2023; 717221, exp. 10/10/2023; 988231, exp. 1/12/2024; 954231, exp. 2/15/2024; m) MMHN37F, UDI/DI none, Lot Numbers: 639221, exp. 12/27/2023; n) MMNU34E, UDI/DI none, Lot Numbers: 928231, exp. 3/13/2024; 889231, exp. 4/21/2024; o) MMUN28E, UDI/DI none, Lot Numbers: 889231, exp. 04/21/2024; p) OSVP16, UDI/DI 00191072156570, Lot Numbers: 827221, exp. 6/17/2023; q) OSVP16, UDI/DI 00191072170835, Lot Numbers: 700221, exp. 10/23/2023; 644221, exp. 12/4/2023; 987231, exp. 11/20/2023; r) SALM40H, UDI/DI none, Lot Numbers: 826221, exp. 6/23/2023; 888221, exp. 4/22/2023; s) SALM40I, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 756221, exp. 9/1/2023; 984231, exp. 1/16/2024; t) SAMB25E, UDI/DI none, Lot Numbers: 843221, exp. 6/6/2023; 894221, exp. 4/16/2023; 871221, exp. 5/9/2023; 686221, exp. 11/10/2023; 758221, exp. 8/30/2023; 793221, exp. 7/26/2023; 974231, exp. 1/26/2024; u) SAMJ36G, UDI/DI none, Lot Numbers: 660221, exp. 12/6/2023; 704221, exp. 10/23/2023; 752221, exp. 9/5/2023; 800221, exp. 7/19/2023; v) SMMN65, UDI/DI 00191072188472, Lot Numbers: 969231, exp. 1/31/2024; 926231, exp. 3/15/2024; w) SMMN65, UDI/DI none, Lot Numbers: 919231, exp. 3/22/2024; x) THGN59D, UDI/DI 00191072191069, Lot Numbers: 805221, exp. 7/14/2023; 721221, exp. 10/6/2023; y) TPMN15, UDI/DI 00191072171771, Lot Numbers: 737221, exp. 9/20/2023; 702221, exp. 10/25/2023; 664221, exp. 12/2/2023; 990231, exp. 1/10/2024; 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; 903231, exp. 4/7/2024; z) TPUN48, UDI/DI 00191072171832, Lot Numbers: 738221, exp. 9/19/2023; 688221, exp. 11/8/2023; 637221, exp. 12/29/2023; 967231, exp. 2/2/2024; 924231, exp. 3/17/2024
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; b) BRACHY THERAPY PROCEDURE PAC, kit number AGBR90E; c) MINI MINOR AGH ASC PACK, kit number AGMM60M; d) OUTPATIENT PORT ACCESS KIT 307907, kit number AHPO55J; e) General Pack, kit number BBGP31C; f) FRACTURE TABLE PACK, kit number FDFT75N; g) Arthroscopy Pack, kit number HCAR04K; h) BARIATRIC PACK, kit number HGBA92F; i) LAPAROTOMY PACK, kit number HNLP12J; j) MIDLINE CHANGE KIT, kit number HSMD20E; k) Minor General, kit number JSMI10I; l) ENDO VASCULAR PACK, kit number LVEV61; m) MAJOR PACK, kit number LVMJ21; n) HEAD AND NECK PACK, kit number MMHN37F; o) KIT, MINOR U BAR, kit number MMNU34E; p) KIT, MAJOR UNIVERSAL, kit number MMUN28E; q) VAG PLASTY PACK, kit number OSVP16; r) DR L MAJOR, kit number SALM40H; s) DR L MAJOR, kit number SALM40I; t) MINOR BASIN SAH, kit number SAMB25E; u) MAJOR SAH, kit number SAMJ36G; v) MINOR PACK, kit number SMMN65; w) General Pack, kit number THGN59D; x) MINOR PACK, kit number TPMN15; y) MAJOR UNIVERSAL PACK, kit number TPUN48 recalled?

Yes. FDA published recall Z-0825-2024 on February 7, 2024. Reported hazard: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail