Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Published March 2, 2016 · Recall initiated: February 4, 2016

Summary

FDA records medical device Class I recall Z-0827-2016, published March 2, 2016 and initiated February 4, 2016. Recalling firm: Draeger Medical, Inc.; Status: Terminated; Reported quantity: 2553; Lot: Catalog #s 8416400 and 8417400. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Draeger Medical, Inc.
Distributor
US Nationwide Distribution including Puerto Rico.
Lot
Catalog #s 8416400 and 8417400
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation. recalled?

Yes. FDA published recall Z-0827-2016 on March 2, 2016. Reported hazard: The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.