This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of human Pax-5 in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using the automated Bond system.
Published March 2, 2016 · Recall initiated: December 30, 2015
Summary
FDA records medical device Class II recall Z-0823-2016, published March 2, 2016 and initiated December 30, 2015. Recalling firm: Leica Microsystems, Inc.; Status: Terminated; Reported quantity: 1700 units; Lot: Product Code/Name: PA0552 (All Lots) Bond Ready-to-Use Primary Antibody PAX-5 (1EW). Detection Kit: DS9800 BondTM Polymer Refine Detection Kit. Affected Detection Kit Lots: 42314, 42388, 42474, 42486, 42499, 42555, 42570, 42579, 42627, 42645, 42683, 42780, 42802, 42823, 42852, 43050. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied after the primary marker incubation as specified in the Instructions for Use Following the Instructions for Use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Leica Microsystems, Inc.
- Distributor
- US Consignees incluude: AR ,AZ ,CA ,CO ,CT ,DC ,DE ,FL ,GA ,HI ,IA ,IL ,KY ,LA ,MA ,MD ,MI ,MN ,MO ,MT ,NC ,NH ,NJ ,NY ,OH ,OK ,PA ,PR ,SC ,TN ,TX ,UT ,VA ,VT ,WA ,WI and WV. Foreign Consignees include:Argentina , Australia , Brazil , Bulgaria , Chile , China , Ecuador ,Egypt , El Salvador , Finland , France , Georgia , Germany , Greece , India , Italy , Japan , Jordan , Kuwait , Lebanon , Malaysia , Philippines , Poland , Portugal , Romania , Russian Fed. , Singapore , South Korea , Spain , Sweden , Switzerland , Thailand , Tunisia , Turkey , United Kingdom and United Arab Emirates.
- Lot
- Product Code/Name: PA0552 (All Lots) Bond Ready-to-Use Primary Antibody PAX-5 (1EW). Detection Kit: DS9800 BondTM Polymer Refine Detection Kit. Affected Detection Kit Lots: 42314, 42388, 42474, 42486, 42499, 42555, 42570, 42579, 42627, 42645, 42683, 42780, 42802, 42823, 42852, 43050.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABOND Ready-To-Use Primary Antibody CDX2 (EP25), REF PA0375September 22, 2021
- FDABOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use.December 18, 2019
- FDALeica M530 OHX Microscope System Supplied with 100 - 120 Vac.August 15, 2018
- FDANussloch GmbH ASP6025 Tissue ProcessorJuly 12, 2017
- FDAGE Precision MPi is an all-digital multipurpose tilt-C x-ray system, intended for a multitude of diagnosticMarch 2, 2016
- FDAEvita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.March 2, 2016
Questions about this recall
Was This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of human Pax-5 in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using the automated Bond system. recalled?
Yes. FDA published recall Z-0823-2016 on March 2, 2016. Reported hazard: The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied after the primary marker incubation as specified in the Instructions for Use Following the Instructions for Use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter