Back to search
FDAFDA · Food and drugs·Recall number Z-0831-2014

Brilliance iCT and Brilliance iCT SP These systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. May include signal analysis and display equipment, patient and equipment supports, components and accessories

Published February 5, 2014 · Recall initiated: July 16, 2013

Summary

FDA records medical device Class II recall Z-0831-2014, published February 5, 2014 and initiated July 16, 2013. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 506; Lot: Model # 728306 100016, 100018, 100019, 100020, 100021, 100022, 100023, 100024, 100025, 100028, 100031, 100032, 100037, 100038, 100039, 100040, 100042, 100043, 100044, 100045, 100046, 100047, 100048, 100049, 100050, 100051, 100052, 100053, 100054, 100055, 100056, 100057, 100058, 100059, 100060, 100061, 100062, 100063, 100064, 100065, 100066, 100067, 100068, 100069, 100070, 100071, 100072, 100073, 100074, 100077, 100078, 100079, 100081, 100082, 100083, 100084, 100085, 100086, 100087, 100088, 100089, 100090, 100091, 100092, 100093, 100094, 100095, 100096, 100097, 100098, 100099, 100100, 100101, 100102, 100103, 100104, 100105, 100107, 100108, 100109, 100110, 100111, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100119, 100120, 100121, 100122, 100123, 100124, 100125, 100126, 100127, 100128, 100129, 100130, 100131, 100132, 100133, 100134, 100135, 100136, 100137, 100138, 100139, 100140, 100141, 100142, 100144, 100145, 100146, 100147, 100148, 100149, 100150, 100151, 100152, 100153, 100154, 100155, 100157, 100158, 100159, 100160, 100161, 100162, 100163, 100164, 100165, 100166, 100167, 100168, 100169, 100170, 100171, 100172, 100173, 100174, 100175, 100176, 100177, 100178, 100179, 100180, 100181, 100182, 100183, 100184, 100185, 100186, 100187, 100188, 100189, 100190, 100191, 100192, 100193, 100194, 100195, 100196, 100197, 100198, 100199, 100200, 100201, 100202, 100203, 100204, 100205, 100206, 100207, 100208, 100209, 100210, 100211, 100212, 100213, 100214, 100215, 100216, 100217, 100218, 100219, 100220, 100221, 100222, 100223, 100224, 100225, 100226, 100227, 100228, 100229, 100230, 100231, 100232, 100233, 100234, 100235, 100236, 100238, 100240, 100241, 100242, 100243, 100244, 100245, 100246, 100247, 100248, 100249, 100250, 100251, 100252, 100253, 100254, 100255, 100256, 100257, 100258, 100259, 100260, 100261, 100262, 100263, 100264, 100265, 100266, 100267, 100268, 100269, 100270, 100271, 100272, 100273, 100274, 100275, 100276, 100277, 100278, 100279, 100280, 100281, 100282, 100283, 100284, 100285, 100286, 100287, 100288, 100289, 100290, 100291, 100292, 100293, 100294, 100295, 100296, 100298, 100299, 100300, 100301, 100302, 100303, 100304, 100305, 100306, 100307, 100308, 100309, 100310, 100311, 100312, 100313, 100314, 100315, 100316, 100317, 100318, 100319, 100320, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100328, 100329, 100330, 100331, 100332, 100333, 100334, 100335, 100336, 100337, 100338, 100339, 100340, 100341, 100342, 100343, 100344, 100345, 100346, 100347, 100348, 100349, 100350, 100351, 100352, 100353, 100354, 100355, 100356, 100357, 100358, 100359, 100360, 100361, 100362, 100363, 100364, 100365, 100366, 100367, 100368, 100369, 100370, 100371, 100372, 100373, 100374, 100375, 100376, 100377, 100378, 100379, 100380, 100381, 100382, 100383, 100384, 100385, 100386, 100387, 100388, 100389, 100391, 100392, 100393, 100394, 100395, 100396, 100397, 100398, 100399, 100400, 100401, 100402, 100403, 100404, 100405, 100406, 100407, 100408, 100409, 100410, 100411, 100412, 100413, 100414, 100415, 100416, 100417, 100418, 100423, 100424, 100426 and 100429. Model # 728311 200003, 200004, 200005, 200006, 200007, 200008, 200009, 200010, 200011, 200012, 200013, 200014, 200015, 200016, 200017, 200018, 200019, 200020, 200021, 200022, 200023, 200024, 200025, 200026, 200027, 200028, 200029, 200030, 200031, 200032, 200033, 200034, 200035, 200036, 200037, 200038, 200039, 200040, 200041, 200042, 200043, 200044, 200045, 200046, 200047, 200048, 200049, 200050, 200051, 200052, 200053, 200054, 200055, 200056, 200057, 200058, 200059, 200060, 200061, 200062, 200063, 200064, 200065, 200066, 200067, 200068, 200069, 200071, 200073, 200074, 200075, 200076, 200077, 200078, 200079, 200080, 200081, 200082, 200083, 200084, 200085, 200086, 200087, 200088, 200089, 200090, 200091, 200092, 200093, 200094, 200095, 200096, 200097, 200098, 200099, 200100, 200101, 200102, 200103, 200104, 200105, 200106, 200107, 200108, 200109, 200110, 200111, 200112, 2001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An artifact that resembles thrombus may appear on the image.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MS, NC, NV, NY, OH, OK, OR, PA, SC, TX, VT, WA, and WV., and the countries of Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brazil, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Netherlands, New Zealand, Norway, Oman, Panama, Philippines, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Uzbekistan.
Lot
Model # 728306 100016, 100018, 100019, 100020, 100021, 100022, 100023, 100024, 100025, 100028, 100031, 100032, 100037, 100038, 100039, 100040, 100042, 100043, 100044, 100045, 100046, 100047, 100048, 100049, 100050, 100051, 100052, 100053, 100054, 100055, 100056, 100057, 100058, 100059, 100060, 100061, 100062, 100063, 100064, 100065, 100066, 100067, 100068, 100069, 100070, 100071, 100072, 100073, 100074, 100077, 100078, 100079, 100081, 100082, 100083, 100084, 100085, 100086, 100087, 100088, 100089, 100090, 100091, 100092, 100093, 100094, 100095, 100096, 100097, 100098, 100099, 100100, 100101, 100102, 100103, 100104, 100105, 100107, 100108, 100109, 100110, 100111, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100119, 100120, 100121, 100122, 100123, 100124, 100125, 100126, 100127, 100128, 100129, 100130, 100131, 100132, 100133, 100134, 100135, 100136, 100137, 100138, 100139, 100140, 100141, 100142, 100144, 100145, 100146, 100147, 100148, 100149, 100150, 100151, 100152, 100153, 100154, 100155, 100157, 100158, 100159, 100160, 100161, 100162, 100163, 100164, 100165, 100166, 100167, 100168, 100169, 100170, 100171, 100172, 100173, 100174, 100175, 100176, 100177, 100178, 100179, 100180, 100181, 100182, 100183, 100184, 100185, 100186, 100187, 100188, 100189, 100190, 100191, 100192, 100193, 100194, 100195, 100196, 100197, 100198, 100199, 100200, 100201, 100202, 100203, 100204, 100205, 100206, 100207, 100208, 100209, 100210, 100211, 100212, 100213, 100214, 100215, 100216, 100217, 100218, 100219, 100220, 100221, 100222, 100223, 100224, 100225, 100226, 100227, 100228, 100229, 100230, 100231, 100232, 100233, 100234, 100235, 100236, 100238, 100240, 100241, 100242, 100243, 100244, 100245, 100246, 100247, 100248, 100249, 100250, 100251, 100252, 100253, 100254, 100255, 100256, 100257, 100258, 100259, 100260, 100261, 100262, 100263, 100264, 100265, 100266, 100267, 100268, 100269, 100270, 100271, 100272, 100273, 100274, 100275, 100276, 100277, 100278, 100279, 100280, 100281, 100282, 100283, 100284, 100285, 100286, 100287, 100288, 100289, 100290, 100291, 100292, 100293, 100294, 100295, 100296, 100298, 100299, 100300, 100301, 100302, 100303, 100304, 100305, 100306, 100307, 100308, 100309, 100310, 100311, 100312, 100313, 100314, 100315, 100316, 100317, 100318, 100319, 100320, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100328, 100329, 100330, 100331, 100332, 100333, 100334, 100335, 100336, 100337, 100338, 100339, 100340, 100341, 100342, 100343, 100344, 100345, 100346, 100347, 100348, 100349, 100350, 100351, 100352, 100353, 100354, 100355, 100356, 100357, 100358, 100359, 100360, 100361, 100362, 100363, 100364, 100365, 100366, 100367, 100368, 100369, 100370, 100371, 100372, 100373, 100374, 100375, 100376, 100377, 100378, 100379, 100380, 100381, 100382, 100383, 100384, 100385, 100386, 100387, 100388, 100389, 100391, 100392, 100393, 100394, 100395, 100396, 100397, 100398, 100399, 100400, 100401, 100402, 100403, 100404, 100405, 100406, 100407, 100408, 100409, 100410, 100411, 100412, 100413, 100414, 100415, 100416, 100417, 100418, 100423, 100424, 100426 and 100429. Model # 728311 200003, 200004, 200005, 200006, 200007, 200008, 200009, 200010, 200011, 200012, 200013, 200014, 200015, 200016, 200017, 200018, 200019, 200020, 200021, 200022, 200023, 200024, 200025, 200026, 200027, 200028, 200029, 200030, 200031, 200032, 200033, 200034, 200035, 200036, 200037, 200038, 200039, 200040, 200041, 200042, 200043, 200044, 200045, 200046, 200047, 200048, 200049, 200050, 200051, 200052, 200053, 200054, 200055, 200056, 200057, 200058, 200059, 200060, 200061, 200062, 200063, 200064, 200065, 200066, 200067, 200068, 200069, 200071, 200073, 200074, 200075, 200076, 200077, 200078, 200079, 200080, 200081, 200082, 200083, 200084, 200085, 200086, 200087, 200088, 200089, 200090, 200091, 200092, 200093, 200094, 200095, 200096, 200097, 200098, 200099, 200100, 200101, 200102, 200103, 200104, 200105, 200106, 200107, 200108, 200109, 200110, 200111, 200112, 2001
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Brilliance iCT and Brilliance iCT SP These systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. May include signal analysis and display equipment, patient and equipment supports, components and accessories recalled?

Yes. FDA published recall Z-0831-2014 on February 5, 2014. Reported hazard: An artifact that resembles thrombus may appear on the image.

What should I do if I have this product?

Voluntary: Firm initiated · Letter