Back to search
FDAFDA · Food and drugs·Recall number Z-0832-2021

Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Published January 20, 2021 · Recall initiated: November 23, 2020

Summary

FDA records medical device Class II recall Z-0832-2021, published January 20, 2021 and initiated November 23, 2020. Recalling firm: Fresenius Medical Care Holdings, Inc.; Status: Ongoing; Reported quantity: 2246 cases; Lot: Lot Numbers: 20LXAC086 20LXAC099 20NXAC014 Additional Affected lots as of 3/17/21: 20LXAC061 20LXAC062 20LXAC099 20LXAC086 20NXAC014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fresenius Medical Care Holdings, Inc.
Distributor
US Nationwide distribution.
Lot
Lot Numbers: 20LXAC086 20LXAC099 20NXAC014 Additional Affected lots as of 3/17/21: 20LXAC061 20LXAC062 20LXAC099 20LXAC086 20NXAC014
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine. recalled?

Yes. FDA published recall Z-0832-2021 on January 20, 2021. Reported hazard: The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

What should I do if I have this product?

Voluntary: Firm initiated · Letter