Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Published January 20, 2021 · Recall initiated: November 23, 2020
Summary
FDA records medical device Class II recall Z-0832-2021, published January 20, 2021 and initiated November 23, 2020. Recalling firm: Fresenius Medical Care Holdings, Inc.; Status: Ongoing; Reported quantity: 2246 cases; Lot: Lot Numbers: 20LXAC086 20LXAC099 20NXAC014 Additional Affected lots as of 3/17/21: 20LXAC061 20LXAC062 20LXAC099 20LXAC086 20NXAC014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Fresenius Medical Care Holdings, Inc.
- Distributor
- US Nationwide distribution.
- Lot
- Lot Numbers: 20LXAC086 20LXAC099 20NXAC014 Additional Affected lots as of 3/17/21: 20LXAC061 20LXAC062 20LXAC099 20LXAC086 20NXAC014
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA5008X Blood Tubing Set, 24 units per case. Includes the below model numbers and descriptions. 1. 5008X HDJuly 15, 2026
- FDABicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;June 17, 2026
- FDABicarby Dialysate; Model number: RFP-400-G;June 17, 2026
- FDADELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for IntraperitonealMay 13, 2026
- FDAMedisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended toJanuary 20, 2021
- FDANaturalyte Liquid Acid concentrate bottles, Product Code 08-2351-8 - Product Usage: used in conjunction withJanuary 20, 2021
Questions about this recall
Was Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine. recalled?
Yes. FDA published recall Z-0832-2021 on January 20, 2021. Reported hazard: The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
What should I do if I have this product?
Voluntary: Firm initiated · Letter