Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended to keep instruments moist during transportation between point of use and cleaning as an initial step of reprocessing.
Published January 20, 2021 · Recall initiated: November 19, 2020
Summary
FDA records medical device Class II recall Z-0830-2021, published January 20, 2021 and initiated November 19, 2020. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 23,445 units, Domestic 5,727 units, Foreign 17,718 units; Lot: Model# M20350 Lot# 1811273,1811277,1906484,1908593,1909635,1910651, 1911658,2001730,2004770 Model# M20358 Lot# 1903408,1906508,1907556,1909644,2001718,2004772 Model# M20359 Lot# 1811278,1906499,1910649,1911680,1911691,1912708, 2001719,2004768 Model# M20400 Lot# 1901340,1901349,1902374,1904448,1904464,1905505, 1906529,1907571,1908595,1909624,1910653,1912700, 2001717,2004766,2004774,2006797. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
STERIS has identified that certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. There is an improbable risk to users of the product from exposure to this bacteria, and no risk to patients.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide distribution - US Nationwide distribution including in the states of CA, DE, CO, VA, FL, WA, TX, OH, PA, MA, IA, CT and the countries of United Kingdom, Saudi Arabia, Portugal, Denmark, Ireland, United Arab Emirates, South Korea, Australia, Greece, Ashmore & Cartier Islands, Belgium, Canada, China, Germany, Japan, Norway, Poland, Qatar, Romania, Sweden, Thailand.
- Lot
- Model# M20350 Lot# 1811273,1811277,1906484,1908593,1909635,1910651, 1911658,2001730,2004770 Model# M20358 Lot# 1903408,1906508,1907556,1909644,2001718,2004772 Model# M20359 Lot# 1811278,1906499,1910649,1911680,1911691,1912708, 2001719,2004768 Model# M20400 Lot# 1901340,1901349,1902374,1904448,1904464,1905505, 1906529,1907571,1908595,1909624,1910653,1912700, 2001717,2004766,2004774,2006797
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDAUroskop Omnia, Model no. 10094910 - Product Usage: a solid state detector fluoroscopic X-ray system, primarilyJanuary 20, 2021
- FDANaturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction withJanuary 20, 2021
Questions about this recall
Was Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended to keep instruments moist during transportation between point of use and cleaning as an initial step of reprocessing. recalled?
Yes. FDA published recall Z-0830-2021 on January 20, 2021. Reported hazard: STERIS has identified that certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. There is an improbable risk to users of the product from exposure to this bacteria, and no risk to patients.
What should I do if I have this product?
Voluntary: Firm initiated · Letter