Back to search
FDAFDA · Food and drugs·Recall number Z-0833-2013

VENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.

Published August 14, 2013 · Recall initiated: September 28, 2012

Summary

FDA records medical device Class II recall Z-0833-2013, published August 14, 2013 and initiated September 28, 2012. Recalling firm: Ventana Medical Systems Inc; Status: Terminated; Reported quantity: 1735 instruments; Lot: Cat#750-600; PN 05342716001; Serial #'s: 310724, 310735, 311187, 312055, 310373, 311052, 311171, 311953, 310750, 311266, 311267, 311564, 311565, 311596, 311304, 310481, 311924, 310607, 311627, 310527, 311436, 311742, 310132, 311432, 310034, 310408, 310687, 310697, 311343, 311958, 310728, 310778, 311991, 312070, 310471, 311326, 311952, 310103, 310777, 311446, 310924, 311303, 311528, 310490, 310913, 311529, 311494, 310139, 310281, 310465, 310275, 310617, 310665, 310716, 311853, 311871, 312137, 312101, 310240, 310237,310238, 311571, 311404, 310356, 310023, 312164, 312165, 310065, 310231, 310711, 310706, 310707, 310703, 311509, 310330, 310633, 311811, 311810, 310977, 311194, 311198, 310208, 310732, 310740, 310187, 311462, 311935, 310145, 311037, 310346, 310585, 310694, 310699, 310701, 310702, 310705, 310709, 310710, 310704, 311106, 311110, 311111, 311419, 311484, 311527, 310345, 311381, 311353, 310480, 311869, 311921, 311458, 310315, 310412, 311988, 310618, 310397, 310399, 311868, 311919, 312139, 312141, 312142, 311270, 310069, 311466, 311929, 310083, 310094, 310154, 311904, 310588, 312098, 312100, 310814, 311731, 311783, 311785, 312096, 312104, 312153, 310227, 310781, 310174, 311403, 311080, 311081, 311121, 311122, 311169, 311188, 311189, 311190, 311242, 311245, 311298, 311926, 310403, 310718, 311416, 311417, 312105, 312112, 310796, 310832, 310824, 311135, 311248, 312066, 310895, 310267, 310272, 311481, 311771, 311772, 310882, 311186, 311937, 310344, 311439, 311505, 310146, 310506, 311545, 310446, 310312, 311584, 312124, 310839, 311319, 311261, 310070, 310508, 311367, 311578, 311576, 311629, 311601, 311600, 311607, 311603, 311602, 311606, 311598, 311597, 311604, 311605, 311599, 311272, 312145, 312123, 310339, 310455, 310472, 310478, 310482, 311562, 311560, 311923, 311498, 311814, 311813, 311804, 310020, 310844, 311313, 311533, 310779, 311241, 311750, 310268, 310331, 311221, 311036, 311034, 311035, 310201, 310321, 310792, 312056, 311147, 311153, 311315, 311308, 310641, 310678, 310878, 310032, 311774, 311182, 310390, 311151, 311268, 311284, 311552, 310902, 311455, 310491, 311038, 312008, 310282, 310696, 311735, 310173, 310402, 311710, 312118, 311874, 310972, 311375, 311727, 312128, 312134, 310933, 310921, 311285, 310297, 310769, 311876, 310916, 311486, 310460, 310245, 310576, 311483, 312097, 312099, 312089, 312109, 311518, 311479, 311901, 312144, 312172, 311551, 310555, 310363, 311808, 310570, 311001, 310358, 310492, 311252, 311687, 311672, 311402, 310114, 310122, 310126, 311300, 310794, 311350, 311592, 310749, 310285, 311299, 310897, 311323, 310258, 310467, 310470, 311628, 310943, 310035, 310066, 310087, 310185, 310752, 310757, 311134, 312138, 310489, 311973, 311191, 310456, 310575, 310673, 310683, 310890, 310821, 312170, 312171, 312060, 312102, 312202, 310500, 310505, 311394, 311917, 311511, 311530, 310894, 312136, 311369, 311665, 312152, 310804, 311517, 311465, 310893, 310896, 311946, 311947, 310323, 310776, 310654, 310299, 310452, 311675, 310404, 310810, 310708, 310712, 311073, 311250, 310910, 311401, 311928, 310941, 310026, 310089, 310336, 310341, 311736, 311610, 311608, 311609, 311728, 310820, 311821, 310141, 310912, 311443, 311442, 311243, 311207, 310256, 310378, 310229, 312163, 312058, 310771, 311103, 310915, 310533, 310124, 311318, 310218, 310811, 311738, 310037, 310038, 310365, 310377, 311210, 312085, 312086, 312087, 311295, 312023, 311493, 311357, 310833, 310800, 311464, 311423, 311173, 311364, 310135, 310608, 310935, 311352, 311354, 311356, 310920, 310183, 310257, 310306, 310852, 311431, 310876, 311949, 311956, 310828, 310830, 310928, 310929, 310930, 310355, 310271, 311384, 312122, 310846, 311345, 310119, 310391, 310881, 310483, 310488, 311893, 312063, 310284, 311663, 312131, 310374, 310507, 310144, 310775, 311730, 311939, 310106, 310205, 310241, 310620, 310713, 310031, 310742, 311016, 311463, 310445, 312117, 310906, 311376, 311377, 311378, 311379, 311380, 310288, 311334, 311957, 311714, 310898, 311477, 310210, 3102. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potiential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform. This type of overflow event can lead to migration of fluid across tubing and wiring within the instrument cabinet resulting in contact with electrical circuitry. This may create shorting conditions and u

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ventana Medical Systems Inc
Distributor
Worldwide distribution: USA (nationwide) and country of: Germany.
Lot
Cat#750-600; PN 05342716001; Serial #'s: 310724, 310735, 311187, 312055, 310373, 311052, 311171, 311953, 310750, 311266, 311267, 311564, 311565, 311596, 311304, 310481, 311924, 310607, 311627, 310527, 311436, 311742, 310132, 311432, 310034, 310408, 310687, 310697, 311343, 311958, 310728, 310778, 311991, 312070, 310471, 311326, 311952, 310103, 310777, 311446, 310924, 311303, 311528, 310490, 310913, 311529, 311494, 310139, 310281, 310465, 310275, 310617, 310665, 310716, 311853, 311871, 312137, 312101, 310240, 310237,310238, 311571, 311404, 310356, 310023, 312164, 312165, 310065, 310231, 310711, 310706, 310707, 310703, 311509, 310330, 310633, 311811, 311810, 310977, 311194, 311198, 310208, 310732, 310740, 310187, 311462, 311935, 310145, 311037, 310346, 310585, 310694, 310699, 310701, 310702, 310705, 310709, 310710, 310704, 311106, 311110, 311111, 311419, 311484, 311527, 310345, 311381, 311353, 310480, 311869, 311921, 311458, 310315, 310412, 311988, 310618, 310397, 310399, 311868, 311919, 312139, 312141, 312142, 311270, 310069, 311466, 311929, 310083, 310094, 310154, 311904, 310588, 312098, 312100, 310814, 311731, 311783, 311785, 312096, 312104, 312153, 310227, 310781, 310174, 311403, 311080, 311081, 311121, 311122, 311169, 311188, 311189, 311190, 311242, 311245, 311298, 311926, 310403, 310718, 311416, 311417, 312105, 312112, 310796, 310832, 310824, 311135, 311248, 312066, 310895, 310267, 310272, 311481, 311771, 311772, 310882, 311186, 311937, 310344, 311439, 311505, 310146, 310506, 311545, 310446, 310312, 311584, 312124, 310839, 311319, 311261, 310070, 310508, 311367, 311578, 311576, 311629, 311601, 311600, 311607, 311603, 311602, 311606, 311598, 311597, 311604, 311605, 311599, 311272, 312145, 312123, 310339, 310455, 310472, 310478, 310482, 311562, 311560, 311923, 311498, 311814, 311813, 311804, 310020, 310844, 311313, 311533, 310779, 311241, 311750, 310268, 310331, 311221, 311036, 311034, 311035, 310201, 310321, 310792, 312056, 311147, 311153, 311315, 311308, 310641, 310678, 310878, 310032, 311774, 311182, 310390, 311151, 311268, 311284, 311552, 310902, 311455, 310491, 311038, 312008, 310282, 310696, 311735, 310173, 310402, 311710, 312118, 311874, 310972, 311375, 311727, 312128, 312134, 310933, 310921, 311285, 310297, 310769, 311876, 310916, 311486, 310460, 310245, 310576, 311483, 312097, 312099, 312089, 312109, 311518, 311479, 311901, 312144, 312172, 311551, 310555, 310363, 311808, 310570, 311001, 310358, 310492, 311252, 311687, 311672, 311402, 310114, 310122, 310126, 311300, 310794, 311350, 311592, 310749, 310285, 311299, 310897, 311323, 310258, 310467, 310470, 311628, 310943, 310035, 310066, 310087, 310185, 310752, 310757, 311134, 312138, 310489, 311973, 311191, 310456, 310575, 310673, 310683, 310890, 310821, 312170, 312171, 312060, 312102, 312202, 310500, 310505, 311394, 311917, 311511, 311530, 310894, 312136, 311369, 311665, 312152, 310804, 311517, 311465, 310893, 310896, 311946, 311947, 310323, 310776, 310654, 310299, 310452, 311675, 310404, 310810, 310708, 310712, 311073, 311250, 310910, 311401, 311928, 310941, 310026, 310089, 310336, 310341, 311736, 311610, 311608, 311609, 311728, 310820, 311821, 310141, 310912, 311443, 311442, 311243, 311207, 310256, 310378, 310229, 312163, 312058, 310771, 311103, 310915, 310533, 310124, 311318, 310218, 310811, 311738, 310037, 310038, 310365, 310377, 311210, 312085, 312086, 312087, 311295, 312023, 311493, 311357, 310833, 310800, 311464, 311423, 311173, 311364, 310135, 310608, 310935, 311352, 311354, 311356, 310920, 310183, 310257, 310306, 310852, 311431, 310876, 311949, 311956, 310828, 310830, 310928, 310929, 310930, 310355, 310271, 311384, 312122, 310846, 311345, 310119, 310391, 310881, 310483, 310488, 311893, 312063, 310284, 311663, 312131, 310374, 310507, 310144, 310775, 311730, 311939, 310106, 310205, 310241, 310620, 310713, 310031, 310742, 311016, 311463, 310445, 312117, 310906, 311376, 311377, 311378, 311379, 311380, 310288, 311334, 311957, 311714, 310898, 311477, 310210, 3102
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents. recalled?

Yes. FDA published recall Z-0833-2013 on August 14, 2013. Reported hazard: Potiential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform. This type of overflow event can lead to migration of fluid across tubing and wiring within the instrument cabinet resulting in contact with electrical circuitry. This may create shorting conditions and u

What should I do if I have this product?

Voluntary: Firm initiated · Letter