DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, REF T653NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Published February 5, 2014 · Recall initiated: September 17, 2013

Summary

FDA records medical device Class II recall Z-0836-2014, published February 5, 2014 and initiated September 17, 2013. Recalling firm: DeRoyal Industries Inc; Status: Terminated; Reported quantity: 409 units; Lot: Lot Numbers: 32901145, 33063999, 3132641, 3217556, 33276468, 33369411, 33412431, 33415413. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
DeRoyal Industries Inc
Distributor
Nationwide Distribution.
Lot
Lot Numbers: 32901145, 33063999, 3132641, 3217556, 33276468, 33369411, 33412431, 33415413
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, REF T653NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy. recalled?

Yes. FDA published recall Z-0836-2014 on February 5, 2014. Reported hazard: The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

What should I do if I have this product?

Voluntary: Firm initiated · Letter