DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee Blanket w/ Straps NS, REF T654NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
Published February 5, 2014 · Recall initiated: September 17, 2013
Summary
FDA records medical device Class II recall Z-0837-2014, published February 5, 2014 and initiated September 17, 2013. Recalling firm: DeRoyal Industries Inc; Status: Terminated; Reported quantity: 48 units; Lot: Lot Numbers: 32871775, 32980891, 33132755, 33141184, 33217572, 33329945. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DeRoyal Industries Inc
- Distributor
- Nationwide Distribution.
- Lot
- Lot Numbers: 32871775, 32980891, 33132755, 33141184, 33217572, 33329945
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAREF: 81-030415 Myocardial Probe Non-Pyrogenic , 400 SeriesSeptember 30, 2026
- FDADeRoyal Industries Inc. Ref 46-406, Covaderm Plus, Vascular Access Dressing 6"x6" (15cmx15cm)June 10, 2026
- FDADeRoyal CHILDREN'S LAPTRACEPACK, 89-9362.06November 13, 2024
- FDADeRoyal C-SECTION TRACEPACK, REF 89-9338-06November 13, 2024
- FDADeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blnkt NS, REF T653NS, Rx only, ManufacturerFebruary 5, 2014
- FDADeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Shoulder Blanket w/ Straps NS, REF T655NS, Rx only,February 5, 2014
Questions about this recall
Was DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee Blanket w/ Straps NS, REF T654NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy. recalled?
Yes. FDA published recall Z-0837-2014 on February 5, 2014. Reported hazard: The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
What should I do if I have this product?
Voluntary: Firm initiated · Letter