Back to search
FDAFDA · Food and drugs·Recall number Z-0839-2019

Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904

Published February 27, 2019 · Recall initiated: December 4, 2018

Summary

FDA records medical device Class II recall Z-0839-2019, published February 27, 2019 and initiated December 4, 2018. Recalling firm: Randox Laboratories Ltd.; Status: Terminated; Reported quantity: 225 kits; Lot: All lot codes. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Randox Laboratories Ltd.
Distributor
WV, PR
UPC
05055273200904
Lot
All lot codes
Risk classification
Class II
Area (FDA)
device
Country
Ireland

Related recalls

Questions about this recall

Was Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904 recalled?

Yes. FDA published recall Z-0839-2019 on February 27, 2019. Reported hazard: Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail

How do I know if my unit is part of the recall?

Compare your product details with the notice. UPC: 05055273200904.