InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908
Published February 5, 2014 · Recall initiated: January 13, 2014
Summary
FDA records medical device Class II recall Z-0845-2014, published February 5, 2014 and initiated January 13, 2014. Recalling firm: Philips Healthcare; Status: Terminated; Reported quantity: 50; Lot: Model Numbers: 861470 (120V), 861472 (240V). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.
What to do
Voluntary: Firm initiated · Other
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare
- Distributor
- Worldwide Distribution - US (Nationwide) and Internationally to: Austria, France, Germany The Netherlands, Spain, Sweden, and Switzerland.
- Lot
- Model Numbers: 861470 (120V), 861472 (240V).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIncisive CT, software version 5.0October 11, 2023
- FDAPhilips Healthcare Brilliance iCT Computed Tomography X-Ray SystemJune 15, 2016
- FDAKODAK DirectView DR 7500 Dual Detector System, MODEL DR 7500, Manufactured by Carestream Health, Inc. Made inFebruary 5, 2014
- FDASimplexa Flu A/B & RSV Direct assay, Model MOL2650. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assayFebruary 5, 2014
- FDAAPI NH (REF 10400) API NH is a standardized system for the identification of Neisseria, Haemophilus (andFebruary 5, 2014
- FDATapered Screw-Vent Implant, MTX, 4.7mm x 11.5mm, 4.5mm Platform. Catalog Number TSVWB11, Lot 62470008. ClassFebruary 5, 2014
Questions about this recall
Was InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908 recalled?
Yes. FDA published recall Z-0845-2014 on February 5, 2014. Reported hazard: Phillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.
What should I do if I have this product?
Voluntary: Firm initiated · Other