Tapered Screw-Vent Implant, MTX, 4.7mm x 11.5mm, 4.5mm Platform. Catalog Number TSVWB11, Lot 62470008. Class II, 510(k) K13227.
Published February 5, 2014 · Recall initiated: January 6, 2014
Summary
FDA records medical device Class II recall Z-0847-2014, published February 5, 2014 and initiated January 6, 2014. Recalling firm: Zimmer Dental Inc; Status: Terminated; Reported quantity: 307; Lot: Catalog Number TSVWB11, Lot 62470008. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Zimmer Dental is conducting a voluntary recall of a single lot of the Tapered Screw-Vent Implant, Catalog TSVWB11, Lot 62470008, because some of the packages of this lot may have the Cap Label state ¿3.7 x 10mm instead of ¿4.7 x 11.5mm.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Dental Inc
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Thailand, Japan, Hong Kong, Morocco, and UAE.
- Lot
- Catalog Number TSVWB11, Lot 62470008.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATapered Screw-Vent MTX Model TSVTB11December 27, 2017
- FDAZimmer Dental Tapered SwissPlus & SwissPlus Implant Systems surgical kit, Ref OPCST. The label shows theDecember 27, 2017
- FDAZimmer Dental GemLock RHD2.5 Hex Driver. The label shows the product is Manufactured by Zimmer Dental Inc.,December 27, 2017
- FDAIMPLANT,1-PC,ANG,3.0X10MM, REF ZOP30A10; IMPLANT, 1-PC, ANG, 3.0X11.5MM, REF ZOP30A11; IMPLANT, 1October 4, 2017
- FDASimplexa Flu A/B & RSV Direct assay, Model MOL2650. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assayFebruary 5, 2014
- FDACompressor Mini To provide a supply of dry, filtered compressed air for a medical respiratory ventilator orFebruary 5, 2014
Questions about this recall
Was Tapered Screw-Vent Implant, MTX, 4.7mm x 11.5mm, 4.5mm Platform. Catalog Number TSVWB11, Lot 62470008. Class II, 510(k) K13227. recalled?
Yes. FDA published recall Z-0847-2014 on February 5, 2014. Reported hazard: Zimmer Dental is conducting a voluntary recall of a single lot of the Tapered Screw-Vent Implant, Catalog TSVWB11, Lot 62470008, because some of the packages of this lot may have the Cap Label state ¿3.7 x 10mm instead of ¿4.7 x 11.5mm.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit