Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.
Published February 7, 2024 · Recall initiated: November 28, 2023
Summary
FDA records medical device Class II recall Z-0846-2024, published February 7, 2024 and initiated November 28, 2023. Recalling firm: XTANT Medical Holdings, Inc; Status: Ongoing; Reported quantity: 6; Lot: UDI-DI: 00810076564005, Lot: AP23000143. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- XTANT Medical Holdings, Inc
- Distributor
- US distribution in MI.
- Lot
- UDI-DI: 00810076564005, Lot: AP23000143
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAXTANT MEDICAL Xpress Pedicle Screw System Dual Lead 6.5 x 50 mm, Part Number X073-6555-DL.April 15, 2026
- FDACortera Spinal Fixation System Screw, Poly, Solid, 6.5mm x 45mm, REF: 150100-65045October 23, 2024
- FDASofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36February 7, 2024
- FDAAzurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)February 7, 2024
- FDASurgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLKFebruary 7, 2024
- FDAAzurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)February 7, 2024
Questions about this recall
Was Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length. recalled?
Yes. FDA published recall Z-0846-2024 on February 7, 2024. Reported hazard: Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
What should I do if I have this product?
Voluntary: Firm initiated · Letter