Back to search
FDAFDA · Food and drugs·Recall number Z-0846-2024

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Published February 7, 2024 · Recall initiated: November 28, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
XTANT Medical Holdings, Inc
Distributor
US distribution in MI.
Lot
UDI-DI: 00810076564005, Lot: AP23000143
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?