Back to search
FDAFDA · Food and drugs·Recall number Z-0870-2024

SCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0028

Published February 7, 2024 · Recall initiated: December 1, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Howmedica Osteonics Corp.
Distributor
Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
Lot
GTIN: 07613327033793 Lot Numbers: A59N, EP7J , X79L
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card