Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts.
Published February 27, 2013 · Recall initiated: October 31, 2012
Summary
FDA records medical device Class II recall Z-0872-2013, published February 27, 2013 and initiated October 31, 2012. Recalling firm: Hill-Rom, Inc.; Status: Terminated; Reported quantity: 3816 in total; Lot: The Liko Overhead Rail System components are not serialized and have been distributed since the early 1980's to the present. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hill-Rom, Inc.
- Distributor
- Worldwide Distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and CANADA LOCATIONS: AB, BC, MB, NB, NS, ON, and SK.
- Lot
- The Liko Overhead Rail System components are not serialized and have been distributed since the early 1980's to the present.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProgressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or otherAugust 5, 2020
- FDASteriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece.July 1, 2015
- FDAViking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.June 24, 2015
- FDAAffinity® Four Birthing Bed, Product Usage: Product is intended to be used as birthing beds for women of childSeptember 10, 2014
- FDAEXPAREL (bupivacaine liposome injectable suspension), 1.3%, 266 mg/20 mL (13.3 mg/mL) single use vials,March 6, 2013
- FDAPhilips HeartStart XL+ Defibrillator/Monitor Model number: 861290 The HeartStart XL+ is intended for use in aFebruary 27, 2013
Questions about this recall
Was Liko Overhead Rail System, Ceiling Mounted Rail Systems for all Liko Stationary Lifts including Likorall, Multirall and Masterlift Systems Product Usage - The Overhead Rail System allows overhead lifts to be mounted to the ceiling allowing versatility for use in lifting situations. It can be used with different Liko ceiling lifts. recalled?
Yes. FDA published recall Z-0872-2013 on February 27, 2013. Reported hazard: Liko/Hill-Rom to date has received two reports from facilities alleging that the overhead rail system has failed to the extent that the overhead rails have fallen down while in use in high humidity (swimming pool) areas. Analysis of the incidents suggest that the components (hardware) used to secure the rail to the ceiling had corroded and may be unsuitable for environments with an atmosphere con
What should I do if I have this product?
Voluntary: Firm initiated · Letter