Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece.
Published July 1, 2015 · Recall initiated: March 2, 2015
Summary
FDA records medical device Class II recall Z-1820-2015, published July 1, 2015 and initiated March 2, 2015. Recalling firm: Hill-Rom, Inc.; Status: Terminated; Reported quantity: 66691 units; Lot: Product #s: 157400; 257400; 157500; 257500. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The metal cannula attached to the handpiece can rotate and become dislodged from the plastic handpiece.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hill-Rom, Inc.
- Distributor
- Worldwide Distribution-US (nationwide) including the states of MI & PA, and the countries of Belgium, England, Italy, Czech Republic, Turkey, Greece, Ireland, Scotland, Spain, and Lebanon.
- Lot
- Product #s: 157400; 257400; 157500; 257500
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProgressa Bed, Catalog No. P7500 - Product Usage: intended to be used to treat or prevent pulmonary or otherAugust 5, 2020
- FDA- Cambridge House River Run Royal Smoked Salmon Products are packaged in Reduced Oxygen Package (ROP). ProductJuly 1, 2015
- FDAFortex" Pedicle Screw System Continuous Radius Rod 5.5 - 50mm, X-Spine Systems, Inc., 452 Alexandersville RoadJuly 1, 2015
- FDAViking M, L, and XL Mobile Lifts. Non-AC-powered patient lift.June 24, 2015
- FDAAffinity® Four Birthing Bed, Product Usage: Product is intended to be used as birthing beds for women of childSeptember 10, 2014
- FDATotalCare Bed used with AccuMax Quantum VPC mattress AccuMax Quantum Complete mattress, and AccuMax QuantumJuly 31, 2013
Questions about this recall
Was Steriseal-Cannula Ophthalmic; Bi-Manual Irrigating/Aspirating Handpiece. recalled?
Yes. FDA published recall Z-1820-2015 on July 1, 2015. Reported hazard: The metal cannula attached to the handpiece can rotate and become dislodged from the plastic handpiece.
What should I do if I have this product?
Voluntary: Firm initiated · Letter