HemoCue® Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252, and 1310257. HemoCue® Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials.
Published December 31, 2014 · Recall initiated: October 20, 2014
Summary
FDA records medical device Class II recall Z-0873-2015, published December 31, 2014 and initiated October 20, 2014. Recalling firm: Radiometer America Inc; Status: Terminated; Reported quantity: 3,712,900 microcuvettes; Lot: Model Number 110705: Lot numbers: 1310500, 1310501, 1310502, 1310503, 1310518, 1310519, 1310520, 1310521, 1310767, 1310768, 1310769, 1310770, 1310771, 1310772, 1310779, 1310780, 1310781, 1310788, 1310789, 1310790, 1310799, 1311524, 1311525, 1311526, 1311527, 1311528, 1311529, 1311530, 1311531, 1311532, 1311533, 1311534, 1311535, 1311536, 1311537, 1311538, 1311539, 1311541, 1311542 Model Number 110706: Lot numbers: 1310207, 1310208, 1310209, 1310210, 1310211, 1310212, 1310213, 1310214, 1310215, 1310216, 1310228, 1310231, 1310232, 1310234, 1310235, 1310238, 1310239, 1310247, 1310249, 1310250, 1310256, 1310257, 1310258, 1310259, 1310260, 1311270, 1311271, 1311272, 1311273, 1311274, 1311279, 1311280, 1311281, 1311282, 1311283, 1311284, 1311291, 1311296, 1311297, 1311314, 1311315, 1311316, 1311317, 1311318, Model Number 110715: Lot numbers: 1310217, 1310218, 1310219, 1310221, 1310222, 1310229, 1310230, 1310237, 1310240, 1310241, 1310242, 1310251, 1310252, 1311267, 1311268, 1311269, 1311298, 1311301, 1311302, 1311303, 1311304, 1311321, 1311322, 1311323, 1311324 Model Number 110716: Lot numbers: 1310202, 1310203, 1310204, 1310223, 1310224, 1310225, 1310227, 1310243, 1310244, 1310245, 1310246, 1310261, 1311263, 1311264, 1311265, 1311266, 1311277, 1311278, 1311305, 1311306, 1311307, 1311308, 1311309, 1311310, 1311311, 1311312, 1311313, 1311325 Model Number 110717: Lot numbers: 1310504, 1310505, 1310506, 1310507, 1310508, 1310509, 1310510, 1310514, 1310515, 1310516, 1310517, 1310773, 1310774, 1310775, 1310776, 1310783, 1310784, 1310785, 1310791, 1310792, 1310794, 1310795, 1310796, 1311540, 1311543, 1311544, 1311545 Model Number 110718: Lot numbers: 1310511, 1310512, 1310513, 1310777, 1310778, 1310786, 1310787, 1310797, 1310798 Model Number 110719: Lot numbers: 1310236, 1310237, 1310248, 1311285, 1311286, 1311287, 1311288, 1311289, 1311290, 1311292, 1311293, 1311294, 1311295, 1311299, 1311300 Model Number 110720: Lot numbers: 1310205, 1310206, 1310226, 1310253, 1310254, 1310255, 1311319, 1311320. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Radiometer America Inc
- Distributor
- Worldwide Distribution-US (nationwide) and the countries of Canada, Austria, AUSTRALIA, SWITZERLAND, COTE D'IVOIRE, CYPRUS, CZECH REPUBLIC, GERMANY, DENMARK, FINLAND, FRANCE, UNITED KINGDOM, GEORGIA, HONG KONG, CROATIA, HUNGARY, IRELAND, INDIA, ICELAND, JAPAN, KENYA, MACEDONIA, MALTA, NETHERLANDS, NORWAY, NEPAL, NEW ZEALAND, POLAND, ROMANIA, RUSSIA, SWEDEN, SLOVENIA, UNITED REPUBLIC OF TANZANIA, and SOUTH AFRICA.
- Lot
- Model Number 110705: Lot numbers: 1310500, 1310501, 1310502, 1310503, 1310518, 1310519, 1310520, 1310521, 1310767, 1310768, 1310769, 1310770, 1310771, 1310772, 1310779, 1310780, 1310781, 1310788, 1310789, 1310790, 1310799, 1311524, 1311525, 1311526, 1311527, 1311528, 1311529, 1311530, 1311531, 1311532, 1311533, 1311534, 1311535, 1311536, 1311537, 1311538, 1311539, 1311541, 1311542 Model Number 110706: Lot numbers: 1310207, 1310208, 1310209, 1310210, 1310211, 1310212, 1310213, 1310214, 1310215, 1310216, 1310228, 1310231, 1310232, 1310234, 1310235, 1310238, 1310239, 1310247, 1310249, 1310250, 1310256, 1310257, 1310258, 1310259, 1310260, 1311270, 1311271, 1311272, 1311273, 1311274, 1311279, 1311280, 1311281, 1311282, 1311283, 1311284, 1311291, 1311296, 1311297, 1311314, 1311315, 1311316, 1311317, 1311318, Model Number 110715: Lot numbers: 1310217, 1310218, 1310219, 1310221, 1310222, 1310229, 1310230, 1310237, 1310240, 1310241, 1310242, 1310251, 1310252, 1311267, 1311268, 1311269, 1311298, 1311301, 1311302, 1311303, 1311304, 1311321, 1311322, 1311323, 1311324 Model Number 110716: Lot numbers: 1310202, 1310203, 1310204, 1310223, 1310224, 1310225, 1310227, 1310243, 1310244, 1310245, 1310246, 1310261, 1311263, 1311264, 1311265, 1311266, 1311277, 1311278, 1311305, 1311306, 1311307, 1311308, 1311309, 1311310, 1311311, 1311312, 1311313, 1311325 Model Number 110717: Lot numbers: 1310504, 1310505, 1310506, 1310507, 1310508, 1310509, 1310510, 1310514, 1310515, 1310516, 1310517, 1310773, 1310774, 1310775, 1310776, 1310783, 1310784, 1310785, 1310791, 1310792, 1310794, 1310795, 1310796, 1311540, 1311543, 1311544, 1311545 Model Number 110718: Lot numbers: 1310511, 1310512, 1310513, 1310777, 1310778, 1310786, 1310787, 1310797, 1310798 Model Number 110719: Lot numbers: 1310236, 1310237, 1310248, 1311285, 1311286, 1311287, 1311288, 1311289, 1311290, 1311292, 1311293, 1311294, 1311295, 1311299, 1311300 Model Number 110720: Lot numbers: 1310205, 1310206, 1310226, 1310253, 1310254, 1310255, 1311319, 1311320
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAABL800 analyzer, model numbers 393-800 and 393-801.January 2, 2019
- FDAAutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+October 11, 2017
- FDAABL90 Flex Model #: 393-090 Portable, automated analyzer that measures pH, blood gases, electrolytes, glucose,March 22, 2017
- FDATCM CombiM Modules 903-111May 11, 2016
- FDAComponent- Dianz Capacitor used in PerfectO2 Oxygen Concentrator, packaged one unit per box.December 31, 2014
- FDASolitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl SleeveDecember 31, 2014
Questions about this recall
Was HemoCue® Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252, and 1310257. HemoCue® Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. recalled?
Yes. FDA published recall Z-0873-2015 on December 31, 2014. Reported hazard: Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag
What should I do if I have this product?
Voluntary: Firm initiated · Letter