QuietCare-Networked product, Facility Server/model: QC101601; The QuietCare system is intended to detect and analyze residents' movements.

Published March 6, 2013 · Recall initiated: December 16, 2012

Summary

FDA records medical device Class II recall Z-0877-2013, published March 6, 2013 and initiated December 16, 2012. Recalling firm: Intel-GE Care Innovations LLC; Status: Terminated; Reported quantity: 19 units; Lot: All units of this model. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Care Innovations detected a software anomaly: During the engineering investigation into the cause of why a Wander Alert was not generated, a defect in the implementation of the Wander and Night Motion ADLs was discovered.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Intel-GE Care Innovations LLC
Distributor
Nationwide Distribution.
Lot
All units of this model.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was QuietCare-Networked product, Facility Server/model: QC101601; The QuietCare system is intended to detect and analyze residents' movements. recalled?

Yes. FDA published recall Z-0877-2013 on March 6, 2013. Reported hazard: Care Innovations detected a software anomaly: During the engineering investigation into the cause of why a Wander Alert was not generated, a defect in the implementation of the Wander and Night Motion ADLs was discovered.

What should I do if I have this product?

Voluntary: Firm initiated · Letter