Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty.

Published March 6, 2013 · Recall initiated: January 15, 2010

Summary

FDA records medical device Class II recall Z-0878-2013, published March 6, 2013 and initiated January 15, 2010. Recalling firm: Medacta Usa Inc; Status: Terminated; Reported quantity: 293 units; Lot: Reference Number Lot Number Reference Number Lot Number 01.26.10.0133: 095920, 096050, 096050A; 01.26.10.0138: 095925, 096055, 096055A; 01.26.10.0134: 095921, 096051, 096051A; 01.26.10.0139: 095926, 096056, 096056A; 01.26.10.0135: 095922, 096052, 096052A; 01.26.10.0140: 095927, 096057, 096057A; 01.26.10.0136: 095923, 096053, 096053A; 01.26.10.0141: 095928, 096058, 096058A; 01.26.10.0137: 095924, 096054, 096054A; 01.26.10.0142: 095929, 096059, 096059A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases. This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to a larger shell size leaving less bone than planned and repositiong/removal of the shell after initial impaction.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medacta Usa Inc
Distributor
Nationwide Distribution including the states of ID, CO, TX, WI, NY, CT, WA, CA, UT, and AZ.
Lot
Reference Number Lot Number Reference Number Lot Number 01.26.10.0133: 095920, 096050, 096050A; 01.26.10.0138: 095925, 096055, 096055A; 01.26.10.0134: 095921, 096051, 096051A; 01.26.10.0139: 095926, 096056, 096056A; 01.26.10.0135: 095922, 096052, 096052A; 01.26.10.0140: 095927, 096057, 096057A; 01.26.10.0136: 095923, 096053, 096053A; 01.26.10.0141: 095928, 096058, 096058A; 01.26.10.0137: 095924, 096054, 096054A; 01.26.10.0142: 095929, 096059, 096059A.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty. recalled?

Yes. FDA published recall Z-0878-2013 on March 6, 2013. Reported hazard: Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases. This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to a larger shell size leaving less bone than planned and repositiong/removal of the shell after initial impaction.

What should I do if I have this product?

Voluntary: Firm initiated · Letter