Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25
Published January 27, 2021 · Recall initiated: December 22, 2020
Summary
FDA records medical device Class II recall Z-0880-2021, published January 27, 2021 and initiated December 22, 2020. Recalling firm: Abbott Point Of Care Inc.; Status: Terminated; Reported quantity: 2,820,000 cartridges US and 2,273,453 catridges OUS; Lot: Stored at room temperature (18 to 30®C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Point Of Care Inc.
- Distributor
- Nationwide
- Lot
- Stored at room temperature (18 to 30®C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAi-STAT EG6+ cartridge; List Number: 03P77-25;May 6, 2026
- FDAi-STAT EG7+ cartridge; List Number: 03P76-25;May 6, 2026
- FDAi-STAT G3+ cartridge; List Number: 03P78-26;May 6, 2026
- FDAi-STAT EG6+ cartridge. List Number: 03P77-25.September 24, 2025
- FDAIMC Wash-Free Hand Sanitizer, (ethyl alcohol), 75%, 1.7 fl. oz. (50 ml) bottles, Distributed by IMC 4284February 3, 2021
- FDAEMBLEM S-ICD Subcutaneous Electrode Model 3501. Component of implantable cardioverter defibrillator (ICD)January 27, 2021
Questions about this recall
Was Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25 recalled?
Yes. FDA published recall Z-0880-2021 on January 27, 2021. Reported hazard: Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail