Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25

Published January 27, 2021 · Recall initiated: December 22, 2020

Summary

FDA records medical device Class II recall Z-0880-2021, published January 27, 2021 and initiated December 22, 2020. Recalling firm: Abbott Point Of Care Inc.; Status: Terminated; Reported quantity: 2,820,000 cartridges US and 2,273,453 catridges OUS; Lot: Stored at room temperature (18 to 30®C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abbott Point Of Care Inc.
Distributor
Nationwide
Lot
Stored at room temperature (18 to 30®C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25 recalled?

Yes. FDA published recall Z-0880-2021 on January 27, 2021. Reported hazard: Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail