Back to search
FDAFDA · Food and drugs·Recall number Z-0880-2021

Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25

Published January 27, 2021 · Recall initiated: December 22, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Abbott Point Of Care Inc.
Distributor
Nationwide
Lot
Stored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)
Risk classification
Class II
Area (FDA)
device
Country
United States