Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260. A component of the Computed Tomography X-Ray System.

Published March 6, 2013 · Recall initiated: October 24, 2012

Summary

FDA records medical device Class II recall Z-0881-2013, published March 6, 2013 and initiated October 24, 2012. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 39 units; Lot: The EBW Serial Numbers subject to software updates are as follows: 8222, 8235, 8270, 8838, 8841, 8862, 9014, 9019, 9127, 10331, 10348, 10402, 10705, 10749, 10827, 11002, 11362, 11365, 11588, 11597, 11652, 11693, 11704, 11724, 11725, 11828, 11857, 11861, 11945, 11949, 11950, 11951, 11952, 11953, 11954, 11955, 11956, 11957, and 11979. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..
Lot
The EBW Serial Numbers subject to software updates are as follows: 8222, 8235, 8270, 8838, 8841, 8862, 9014, 9019, 9127, 10331, 10348, 10402, 10705, 10749, 10827, 11002, 11362, 11365, 11588, 11597, 11652, 11693, 11704, 11724, 11725, 11828, 11857, 11861, 11945, 11949, 11950, 11951, 11952, 11953, 11954, 11955, 11956, 11957, and 11979.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260. A component of the Computed Tomography X-Ray System. recalled?

Yes. FDA published recall Z-0881-2013 on March 6, 2013. Reported hazard: Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

What should I do if I have this product?

Voluntary: Firm initiated · Letter