Back to search
FDAFDA · Food and drugs·Recall number Z-0890-2013

Atrapro Antipruritic Hydrogel, 1.5 oz gel tube, 4 Net wt/ 1.5 oz.. RX only, Professional Sample, Not for Sale. Pre-printed tube: MS31071, Shelf carton: PL51091-08 Atrapro Antipruritic HydroGel is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including radiation dermatitis and atopic dermatitis.

Published March 6, 2013 · Recall initiated: February 12, 2013

Summary

FDA records medical device Class II recall Z-0890-2013, published March 6, 2013 and initiated February 12, 2013. Recalling firm: Oculus Innovative Sciences Inc; Status: Terminated; Reported quantity: 56,566 1.5 oz tubes; Lot: Lot 12A0012, Exp 2/2013; Lot 12A0013, Exp 2/2013: Lot 12A1116, Exp 2.2013; Lot 12A1119, Exp 2/2013; Lot 12B1144, Exp 3/2013; Lot 12B1148, Exp 3/2013; Lot 12C1157, Exp 4/2013; Lot 12E190. Exp 6/2013; Lot 12E194, Exp 6/2013; Lot 12E198, Exp 6/2013; Lot 12E202. Exp 6/2013; Lot 12E208, Exp 6/2013; Lot 12E211, Exp 6/2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Out of specification result - stability test failure of free available chlorine (FAC) which functions as a preservative for the hydrogel. The 13 month test result obtained was 4 ppm FAC and the stability specification is >10 ppm FAC.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Oculus Innovative Sciences Inc
Distributor
Nationwide Distribution - One distributor in PA
Lot
Lot 12A0012, Exp 2/2013; Lot 12A0013, Exp 2/2013: Lot 12A1116, Exp 2.2013; Lot 12A1119, Exp 2/2013; Lot 12B1144, Exp 3/2013; Lot 12B1148, Exp 3/2013; Lot 12C1157, Exp 4/2013; Lot 12E190. Exp 6/2013; Lot 12E194, Exp 6/2013; Lot 12E198, Exp 6/2013; Lot 12E202. Exp 6/2013; Lot 12E208, Exp 6/2013; Lot 12E211, Exp 6/2013.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Atrapro Antipruritic Hydrogel, 1.5 oz gel tube, 4 Net wt/ 1.5 oz.. RX only, Professional Sample, Not for Sale. Pre-printed tube: MS31071, Shelf carton: PL51091-08 Atrapro Antipruritic HydroGel is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including radiation dermatitis and atopic dermatitis. recalled?

Yes. FDA published recall Z-0890-2013 on March 6, 2013. Reported hazard: Out of specification result - stability test failure of free available chlorine (FAC) which functions as a preservative for the hydrogel. The 13 month test result obtained was 4 ppm FAC and the stability specification is >10 ppm FAC.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail