Back to search
FDAFDA · Food and drugs·Recall number Z-0895-2024

SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508

Published February 7, 2024 · Recall initiated: December 1, 2023

Summary

FDA records medical device Class II recall Z-0895-2024, published February 7, 2024 and initiated December 1, 2023. Recalling firm: Howmedica Osteonics Corp.; Status: Ongoing; Reported quantity: 137 units US; 326 units OUS; Lot: GTIN: 07613327033885 Lot Numbers: EPXH, DYAE, XVK2, PVP9, LJ0K, 12R3, N8P6, T355, 8HDN, A3EV, 5LKW, R04Y, RKT1,D06K, 01JR, 3PA4, 0AYL, W424, 53T1, Y7X2,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Howmedica Osteonics Corp.
Distributor
Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
Lot
GTIN: 07613327033885 Lot Numbers: EPXH, DYAE, XVK2, PVP9, LJ0K, 12R3, N8P6, T355, 8HDN, A3EV, 5LKW, R04Y, RKT1,D06K, 01JR, 3PA4, 0AYL, W424, 53T1, Y7X2,
Risk classification
Class II
Area (FDA)
device
Country
United States