HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-28J
Published February 7, 2024 · Recall initiated: December 1, 2023
Summary
FDA records medical device Class II recall Z-0901-2024, published February 7, 2024 and initiated December 1, 2023. Recalling firm: Howmedica Osteonics Corp.; Status: Ongoing; Lot: GTIN: n/a Lot Numbers: 38059601. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Howmedica Osteonics Corp.
- Distributor
- Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.
- Lot
- GTIN: n/a Lot Numbers: 38059601
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.October 21, 2026
- FDATriathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.October 21, 2026
- FDATS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.October 21, 2026
- FDATS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.October 21, 2026
- FDAScorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0910February 7, 2024
- FDAHOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32JFebruary 7, 2024
Questions about this recall
Was HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-28J recalled?
Yes. FDA published recall Z-0901-2024 on February 7, 2024. Reported hazard: Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
What should I do if I have this product?
Voluntary: Firm initiated · Letter