Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes such as Glucose, Magnesium, and Potassium.
Published March 13, 2013 · Recall initiated: November 28, 2012
Summary
FDA records medical device Class II recall Z-0913-2013, published March 13, 2013 and initiated November 28, 2012. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 28 in U.S.; 4 Canada; 318 to other foreign countries.; Lot: All serial numbers and all software versions 4.05 and prior. Part# Item# A94907 AU5811-02 A94912 AU5821-02 A94917 AU5831-02 A94922 AU5841-02 A94908 AU5811-03 A94913 AU5821-03 A94918 AU5831-03 A94923 AU5841-03 A94909 AU5811-04 A94914 AU5821-04 A94919 AU5831-04 A94924 AU5841-04 A94910 AU5811-06 A94915 AU5821-06 A94920 AU5831-06 A94925 AU5841-06. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL result.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Austria, Brazil, China, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Korea Republic, Malaysia, Netherlands, Russian Federation, Spain, Switzerland, Taiwan, Turkey, United Kingdom, and Viet Nam.
- Lot
- All serial numbers and all software versions 4.05 and prior. Part# Item# A94907 AU5811-02 A94912 AU5821-02 A94917 AU5831-02 A94922 AU5841-02 A94908 AU5811-03 A94913 AU5821-03 A94918 AU5831-03 A94923 AU5841-03 A94909 AU5811-04 A94914 AU5821-04 A94919 AU5831-04 A94924 AU5841-04 A94910 AU5811-06 A94915 AU5821-06 A94920 AU5831-06 A94925 AU5841-06
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDAAtricure Inc., Max1 Isolator Transpolar Pen. Model A000163. Used to ablate cardiac tissues during cardiacMarch 13, 2013
- FDALifeguard Infusion Set with Smartsite, 20Ga, 1, CLGY-2010 Huber type needle set used for the infusion ofMarch 13, 2013
Questions about this recall
Was Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes such as Glucose, Magnesium, and Potassium. recalled?
Yes. FDA published recall Z-0913-2013 on March 13, 2013. Reported hazard: Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL result.
What should I do if I have this product?
Voluntary: Firm initiated · Letter