Back to search
FDAFDA · Food and drugs·Recall number Z-0915-2013

AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.

Published March 13, 2013 · Recall initiated: February 12, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.

What to do

Voluntary: Firm initiated · Visit

Recall details

Manufacturer
Steris Corporation
Distributor
Worldwide Distribution - USA including Canada and the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, SK, TN, TX, UT, VA, VT, WA, WI, and WV.
Lot
AMSCO C: Serial Numbers: 021911201  023461202 AMSCO 400: Serial Numbers: 030181215-030291310
Risk classification
Class II
Area (FDA)
device
Country
United States