AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
Published March 13, 2013 · Recall initiated: February 12, 2013
Summary
FDA records medical device Class II recall Z-0915-2013, published March 13, 2013 and initiated February 12, 2013. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 395 Units; Lot: AMSCO C: Serial Numbers: 021911201 023461202 AMSCO 400: Serial Numbers: 030181215-030291310. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
What to do
Voluntary: Firm initiated · Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide Distribution - USA including Canada and the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, SK, TN, TX, UT, VA, VT, WA, WI, and WV.
- Lot
- AMSCO C: Serial Numbers: 021911201 023461202 AMSCO 400: Serial Numbers: 030181215-030291310
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDALifeguard Infusion Set with Smartsite, 20Ga, 1, CLGY-2010 Huber type needle set used for the infusion ofMarch 13, 2013
- FDAD3000 Advantage Drive System, 115v Product Usage: The Advantage Drive Systems function as a powered instrumentMarch 13, 2013
Questions about this recall
Was AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities. recalled?
Yes. FDA published recall Z-0915-2013 on March 13, 2013. Reported hazard: The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
What should I do if I have this product?
Voluntary: Firm initiated · Visit