B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.
Published February 12, 2014 · Recall initiated: December 13, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- B. Braun Medical, Inc.
- Distributor
- US Distribution one importer/wholesale distributor in MO.
- Lot
- Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States