B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.
Published February 12, 2014 · Recall initiated: December 13, 2013
Summary
FDA records medical device Class II recall Z-0921-2014, published February 12, 2014 and initiated December 13, 2013. Recalling firm: B. Braun Medical, Inc.; Status: Terminated; Reported quantity: 30, 240 units; Lot: Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B. Braun Medical, Inc.
- Distributor
- US Distribution one importer/wholesale distributor in MO.
- Lot
- Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock,September 25, 2024
- FDAInfusomat UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV-Used with an electrically-powered infusion pump forJuly 24, 2024
- FDAInfusomat UNIV. 15 DROP PUMP SET W/ 0.2 FILTER- Used with an electrically-powered infusion pump forJuly 24, 2024
- FDAInfusomat SPACE PUMP IV SET, 127 IN.-Used with an electrically-powered infusion pump for intravenous (IV)July 24, 2024
- FDAFresenius Bicarbonate Jug Adapter to the 2008 Series Hemodialysis machine Part Number: 650103. The adapter isFebruary 12, 2014
- FDASuperElastic Regular force NiTi lower or upper Natural arches. 0.16 size 10 or 100 count packages. HighlandFebruary 12, 2014
Questions about this recall
Was B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications. recalled?
Yes. FDA published recall Z-0921-2014 on February 12, 2014. Reported hazard: Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.
What should I do if I have this product?
Voluntary: Firm initiated · Letter