ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))

Published February 5, 2020 · Recall initiated: December 9, 2019

Summary

FDA records medical device Class II recall Z-0921-2020, published February 5, 2020 and initiated December 9, 2019. Recalling firm: Mindray DS USA, Inc. dba Mindray North America; Status: Terminated; Reported quantity: 6; Lot: Serial Numbers W2-6C000001, W2-6C000002, W2-6C000003, W2-6C000004, W2-6C000005, XP-5B000611T. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Mindray DS USA, Inc. dba Mindray North America
Distributor
The products were distributed to the following US states: MI and TX.
Lot
Serial Numbers W2-6C000001, W2-6C000002, W2-6C000003, W2-6C000004, W2-6C000005, XP-5B000611T
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring)) recalled?

Yes. FDA published recall Z-0921-2020 on February 5, 2020. Reported hazard: The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.

What should I do if I have this product?

Voluntary: Firm initiated · Letter