SuperElastic Regular force NiTi lower or upper Natural arches. 0.14 size 10 or 100 count packages. Highland Metals Inc. San Jose, CA The devices are intended to provide force to the teeth to effect movement in order to alter their position.
Published February 12, 2014 · Recall initiated: December 13, 2013
Summary
FDA records medical device Class II recall Z-0923-2014, published February 12, 2014 and initiated December 13, 2013. Recalling firm: Highland Metals, Inc.; Status: Terminated; Reported quantity: 37 packages; Lot: Part Number Part Description Lot Number Label Form Actual Form 11100141124 SE Niti 014 Nat U 100 41756 Upper Lower 11100141144 SE Niti 014 Nat L 100 41760 Lower Upper 11100141145 SE Niti 014 Nat L 10 41760 Lower Upper. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Labeling issue: Product labeled as "Upper" may contain "Lower" archwires with a red midline mark. Product labeled ad "Lower" may contain "Upper" archwires with a black midline mark.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Highland Metals, Inc.
- Distributor
- Worldwide Distribution - USA including IL, KY, MI, NY, CA, FL, LA and Internationally to Switzerland., Denmark, Greece, Australia, England, and Canada.
- Lot
- Part Number Part Description Lot Number Label Form Actual Form 11100141124 SE Niti 014 Nat U 100 41756 Upper Lower 11100141144 SE Niti 014 Nat L 100 41760 Lower Upper 11100141145 SE Niti 014 Nat L 10 41760 Lower Upper
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASuperElastic Regular force NiTi lower or upper Natural arches. 0.16 size 10 or 100 count packages. HighlandFebruary 12, 2014
- FDAABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality controlFebruary 12, 2014
- FDAB. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and localFebruary 12, 2014
- FDASmartSite Needleless Connector, Model No. 2000E The SmartSite Needleless Connectors are sterile, individuallyFebruary 12, 2014
- FDAFresenius Bicarbonate Jug Adapter to the 2008 Series Hemodialysis machine Part Number: 650103. The adapter isFebruary 12, 2014
- FDAThe Rotoprone therapy System is an advanced patient care system for the treatment and prevention of pulmonaryFebruary 12, 2014
Questions about this recall
Was SuperElastic Regular force NiTi lower or upper Natural arches. 0.14 size 10 or 100 count packages. Highland Metals Inc. San Jose, CA The devices are intended to provide force to the teeth to effect movement in order to alter their position. recalled?
Yes. FDA published recall Z-0923-2014 on February 12, 2014. Reported hazard: Labeling issue: Product labeled as "Upper" may contain "Lower" archwires with a red midline mark. Product labeled ad "Lower" may contain "Upper" archwires with a black midline mark.
What should I do if I have this product?
Voluntary: Firm initiated · Letter